FDA Adverse Event Malfunction Summary report: N

INSET GUARD

MDR report key: 24701576 · Received March 26, 2026

Report

Report Number
8021545-2026-99989
Event Type
Malfunction
Date Received
March 26, 2026
Date of Event
February 5, 2024
Report Date
February 25, 2026
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244022683
PMA / PMN Number
K210544
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4) DEVICE 1 OF 2. E1: PATIENT CITY: (B)(6) PATIENT COUNTRY: FINLAND. UNOMEDICAL HEREBY SUBMITS THIS INITIAL REPORT AS PART OF ITS COMPLAINT REMEDIATION ACTIVITIES CONDUCTED UNDER A CORRECTIVE AND PREVENTIVE ACTION (CAPA)/FDA ACTION PLAN. THIS SUBMISSION INCLUDES A RETROSPECTIVE REASSESSMENT OF PREVIOUSLY EVALUATED COMPLAINTS. UNOMEDICAL IS PROVIDING THIS INITIAL INFORMATION IN ACCORDANCE WITH THE REPORTING REQUIREMENTS SET FORTH IN 21 CFR PART 803. ANY FIELDS LEFT BLANK INDICATE THAT THE INFORMATION IS UNKNOWN, UNAVAILABLE, OR REMAINS UNCHANGED BASED ON THE REVIEW COMPLETED ON 25-FEB-2026 AND INVESTIGATION COMPLETED ON 16-MAR-2026 THIS MEDICAL DEVICE RECORD (MDR) IS BEING SUBMITTED AS THE INITIAL REPORT. ADDITIONAL INFORMATION - THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS REVIEW OF COMPLAINT RECORD DATABASE 1825026 EVENT DESCRIPTION AND ALL AVAILABLE INFORMATION WAS COMPLETED, AND LOT NUMBER 6002088 WAS PROVIDED. COMPLAINT WAS CLASSIFIED UNDER MALFUNCTION CODE: INFUSION SITE EGRESS Â - TRACE MOISTURE / SUPERFICIAL WETNESS. A QUERY WAS RUN IN THE ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS) ON 16-MAR-2026 AGAINST LOT NUMBER 6002088 AND SIMILAR MALFUNCTION CODE(S). INFUSION SITE EGRESS Â - TRACE MOISTURE / SUPERFICIAL WETNESS INFUSION SITE LEAKAGE - TRACE MOISTURE / SUPERFICAL WETNESS INFUSION SITE LEAKAGE Â - TRACE MOISTURE / SUPERFICAL WETNESS LEAKAGE INFUSION SITE (CANNULA BASE PART/CANNULA) (SPECIFIC CAUSE NOT IDENTIFIED) LEAKAGE FROM INFUSION SITE (CANNULA BASE PART/CANNULA) (SPECIFIC CAUSE NOT IDENTIFIED) NO RECORDS WERE IDENTIFIED FOR THIS LOT AND SIMILAR RELATED ISSUES. A NON-CONFORMANCE (NC)/CORRECTIVE AND PREVENTIVE ACTION (CAPA) QUERY SEARCH WAS RUN IN THE EQMS SYSTEM PERFORMED ON 16-MAR-2026 AGAINST LOT/BATCH NUMBER 6002088. NO RECORDS WERE FOUND. DEVICE HISTORY RECORD (DHR) REVIEW: THE DEVICE HISTORY RECORD (DHR) FOR LOT 6002088 WAS REVIEWED. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE . CONCLUSION OF COMPLAINT INVESTIGATION: LOT NO. 6002088 WAS PROVIDED, HOWEVER, AS THE COMPLAINT IS CATEGORIZED AS A REPORTABLE WITH NO HARM REPORTED AND NO ISSUES WERE FOUND DURING EQMS SEARCH AND DHR REVIEW: NO VISUAL INSPECTION IS REQUIRED TO BE PERFORMED. NO RETAIN SAMPLE TESTING IS REQUIRED TO BE CONDUCTED. NO FURTHER ASSESSMENT WILL BE MADE REGARDING POTENTIAL PRODUCT PERFORMANCE ISSUES, COMPONENT INTEGRITY, OR MANUFACTURING DEFECTS. CAPA DETERMINATION: BASED ON THE AVAILABLE INFORMATION, NO FURTHER INVESTIGATION IS REQUIRED NOR CAPA PLAN IS NEEDED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING (MONTHLY TRIPS AND ALERTS). SUMMARY CONCLUSION: BASED ON THE AVAILABLE INFORMATION, THE INVESTIGATION HAS BEEN PERFORMED, AND THE COMPLAINT COULD NOT BE CONFIRMED AS ANALYZED. THEREFORE, THE COMPLAINT IS CLOSED BASED ON THE EVENT DESCRIPTION AND ALL INFORMATION MADE AVAILABLE.

Description of Event or Problem · 0

THIS EMDR IS BEING SUBMITTED FOR AN UNKNOWN NUMBER QUANTITY. REFERENCE NUMBER (B)(4) EVENT OCCURED IN FINLAND. IT WAS REPORTED THAT THE PATIENT EXPERIENCED LEAKAGE AT THE INFUSION SITE ON (B)(6) 2024. THE INFUSION SET HAD BEEN IN USE FOR 5 DAYS. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
766501 INSET GUARD UNO EWIS BLUE 60/6 HCAP 10PK INT FPA UNOMEDICAL A/S MMT-431A 6002088 05705244022683

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown