FDA Adverse Event Malfunction Summary report: N

DISCOVERY XR650

MDR report key: 2469944 · Received February 20, 2012

Report

Report Number
2126677-2012-00026
Event Type
Malfunction
Date Received
February 20, 2012
Date of Event
January 24, 2012
Report Date
January 24, 2012
Manufacturer
GE MEDICAL SYSTEMS, LLC.
Product Code
KPR
PMA / PMN Number
K051967
Removal / Correction Number
2126677-10/5/11-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS MALFUNCTION WAS DETERMINED TO BE REPORTABLE AS THIS SAME MALFUNCTION HAS PREVIOUSLY CONTRIBUTED TO A SERIOUS INJURY WITHIN THE LAST TWO YEARS. REFERENCE MDR 2126677-2010-00010. GE HEALTHCARE'S INVESTIGATION CONCLUDED THAT THE MECHANISM WHICH APPLIES FRICTION TO THE DISCOVERY XR650 PT BARRIER PIVOTING ARM WEARS RAPIDLY ALLOWING THE ARM TO SWING FREELY WHEN THE LOCK IS NOT APPLIED. THE RESULT IS THAT THE PIVOTING ARM MAY BE STORED IN AN UPRIGHT POSITION WHICH WILL BE UNSTABLE DUE TO A LACK OF RESIDUAL FRICTION. GE HEALTHCARE WILL MODIFY THE ROTATING ARMS TO RESTORE THE RESIDUAL FRICTION SO THAT THE ARM WILL RETAIN POSITION (RATHER THAN SWING FREELY) WHILE THE LOCKING LEVERS ARE DISENGAGED. THE PRODUCT LABELING WILL ALSO BE UPDATED TO IMPROVE AWARENESS AND IDENTIFY THE HAZARD RELATIVE TO PROPER STORAGE. THIS ACTION HAS BEEN REPORTED TO FDA PER 21 CFR PART 806 ON 10/05/2011. REFERENCE CORRECTIONS AND REMOVALS REPORT NUMBER 2126677-10/5/11-002-C.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PT BARRIER ROTATIONAL HANDLE LOCK WAS BROKEN IN A MANNER THAT PREVENTED IT FROM LOCKING INTO POSITION. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DISCOVERY XR650 STATIONARY X-RAY SYSTEM KPR GE MEDICAL SYSTEMS, LLC.

Patients

Seq Age Sex Outcome Treatment
1