SYNCHRON® LX20 PRO ANALYZER
Report
- Report Number
- 2050012-2012-00581
- Event Type
- Malfunction
- Date Received
- February 27, 2012
- Date of Event
- February 3, 2012
- Report Date
- February 6, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K011213
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
BEC HOTLINE HAD THE CUSTOMER TIGHTEN THE SAMPLE DETECTOR AND CLEAN THE CUP, PRIME THE MODULE TO CHECK FOR FILLING/DRAINING AND BUBBLES. THE CUSTOMER WAS ALSO INSTRUCTED TO PERFORM A LAMP CALIBRATION. NO FURTHER ISSUES WERE NOTED. A CLEAR ROOT CAUSE FOR THIS EVENT COULD NOT BE IDENTIFIED.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REGARDING ONE HIGH PHOSPHOROUS (PHOS) PATIENT RESULT OF 9.6MG/DL GENERATED BY THEIR SYNCHRON LX20 PRO ANALYZER. BEC FIELD SERVICE ENGINEER (FSE) WAS ON-SITE A WEEK BEFORE THE EVENT DUE TO A NOISE ERROR. PER FSE, THE PROBLEM WAS A LOOSE MODULE. FSE TIGHTENED THE MODULE. THE MORNING OF THE EVENT, CUSTOMER WAS HAVING PROBLEMS WITH NOISE AND FAILING RANGE WHEN TRYING TO CALIBRATE. THE CUSTOMER WAS USING PHOS REAGENT LOT# M103025 THAT WAS LOADED ON (B)(6) 2012. THE CALIBRATION PASSED ON THE THIRD CALIBRATION, AND THE QC PASSED. THE CUSTOMER THEN RAN SAMPLES AND GOT THE CRITICAL HIGH 9.6MG/DL PHOS RESULT. THE CUSTOMER STOPPED RUNNING PHOS ON THIS INSTRUMENT. THE SAMPLE WAS REPEATED ON A DIFFERENT INSTRUMENT AND GAVE A RESULT OF 3.2MG/DL. THE REAGENT IN USE ON THIS INSTRUMENT WAS LOT# M108199. THE ERRONEOUS RESULT WAS NOT REPORTED OUTSIDE THE LABORATORY AND THERE WAS NO AFFECT TO PATIENT WITH REGARD TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® LX20 PRO ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |