FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE", 10 MM, 30°, PAL, HD COMPATIBLE, AUTOCLAVABLE

MDR report key: 24694350 · Received March 25, 2026

Report

Report Number
9610773-2026-02270
Event Type
Malfunction
Date Received
March 25, 2026
Date of Event
January 21, 2026
Report Date
March 25, 2026
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
NWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE ADDITIONAL MALFUNCTION DAMAGED FIBRE BONDING IN THE OUTER TUBE AREA (DISTAL END) WAS TRACED TO COMPONENT FAILURE, AND PROBABLE CAUSE OF FOREIGN MATERIAL IN LIGHT GUIDE CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION COULD NOT BE ESTABLISHED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE EXHIBITED DAMAGED FIBRE BONDING IN THE OUTER TUBE AREA (DISTAL END), AND FOREIGN MATERIAL IN LIGHT GUIDE CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
745345 VIDEO TELESCOPE "ENDOEYE", 10 MM, 30°, PAL, HD COMPATIBLE, AUTOCLAVABLE RIGID VIDEO LAPAROSCOPE NWB OLYMPUS WINTER & IBE GMBH WA50012A

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown