INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2026-05368
- Event Type
- Injury
- Date Received
- March 25, 2026
- Date of Event
- January 13, 2026
- Report Date
- March 25, 2026
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT CAPSULAR CONTRACTURE WAS RECEIVED ON APRIL 27, 2026, WITH LOT NUMBER 1213488. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: ¿ CAPSULAR CONTRACTURE: UNABLE TO OBSERVE THROUGH VISUAL INSPECTION AS IT IS A PHYSIOLOGICAL PHENOMENON. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (DEFORMATION AND CREASES) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE BAKER GRADE IV.
HEALTHCARE PROFESSIONAL REPORTED "GRADE IV CAPSULAR CONTRACTURE". PATIENT WAS PRESCRIBED WITH SINGULAIR. THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED "GRADE IV CAPSULAR CONTRACTURE". PATIENT WAS PRESCRIBED WITH SINGULAIR. THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 756228 | INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 1213488 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Female | Required Intervention |