DAVINCI 5
Report
- Report Number
- 2955842-2026-18499
- Event Type
- Malfunction
- Date Received
- March 25, 2026
- Date of Event
- April 3, 2025
- Report Date
- March 25, 2026
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874119662
- PMA / PMN Number
- K232610
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- 003
Narratives
AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE REPLACED THE COMMON COMPUTE CONTROLLER (CCC) BOARD AT THE VISION SIDE CART (VSC). FSE CLEANED ALL CONNECTION FIBER POINTS TO THE VARIOUS SUB SYSTEMS FROM THE TOWER. SYSTEM WAS NOW COMMUNICATING WITH ALL SUB SYSTEMS, AND POWERS ON NORMALLY. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS (FA). THE CCC WAS ANALYZED AND THE ERRORS 170 AND 41113 WERE FOUND SEVERAL TIME INDICATING THAT A GIVEN FEATURE MONITORED BY THE FEATURE HEALTH MONITOR IS IN INVALID STATE AND A SUPERVISOR CONSOLE CAPSULE DETECTS THE CORE IS OFFLINE, RESPECTIVELY. UPON VISUAL INSPECTION, NO ISSUES WERE FOUND THAT WOULD BE RELATED TO THE REPORTED EVENT. THIS VSC CCC WAS INSTALLED, AND TESTED ON THE DV5 IN HOUSE SYSTEM WHERE PROGRAMMING WAS NOT ABLE TO PROGRAM, VSC CART TOWER WAS NOT STARTING UP IN BOTH MAINTENANCE MODE AND IN NORMAL MODE, VSC CART POWER BUTTON CONTINUE TO FLASH IN CIRCLE NONSTOP INDICATING STARTUP FAILURE, THEREFORE REPLICATING THE REPORTED EVENT. THE COMPLAINT WAS CONFIRMED BASED ON FAILURE ANALYSIS, WHICH INDICATES THAT THE DEVICE MAY HAVE CONTRIBUTED TO THE CUSTOMER REPORTED ISSUE.
IT WAS REPORTED THAT THE CLINICAL SALES REPRESENTATIVE (CSR) CONTACTED THE TECHNICAL SUPPORT ENGINEER (TSE) TO REPORT THAT ERRORS OCCURRED ON THE SYSTEM. WHEN POWERING ON AT THE TOWER, ALL POWER BUTTONS SPIN BLUE, BUT THE SYSTEM NEVER COMPLETES SELF-TEST. LOGS SHOW POSSIBLE PROBLEM WITH CONSOLE2 SIM. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 539772 | DAVINCI 5 | VISION SIDE CART | NAY | INTUITIVE SURGICAL, INC | 380746-40 | N/A | 00886874119662 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |