FDA Adverse Event Other Summary report: N

MEDCARE

MDR report key: 2469298 · Received February 15, 2012

Report

Report Number
2135150-2012-00003
Event Type
Other
Date Received
February 15, 2012
Date of Event
January 24, 2012
Report Date
February 15, 2012
Manufacturer
MEDCARE PRODUCTS
Product Code
FNG
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

A REP FROM MEDCARE WENT TO THE FACILITY TO INVESTIGATE. IT WAS DETERMINED THAT THE APPROXIMATE AGE OF THE BELT IN USE WAS 7 YRS. BASED ON THE CONDITION IT WAS IN IT SHOULD HAVE BEEN REMOVED FROM SERVICE. SLINGS AND BELTS ARE TO INSPECTED FOR WEAR AND TEAR BEFORE AND AFTER EACH USE. ALSO, IT IS HARD TO UNDERSTAND THE CLAIM THAT THE RESIDENT FELL BACKWARDS BECAUSE THE BELT CAME UNCLASPED. THE BELT IS FASTENED IN THE FRONT OF THE RESIDENT. IF THIS COMES UNCLASPED THE RESIDENT'S BACKSIDE WOULD STILL BE SUPPORTED.

Description of Event or Problem · 1

RESIDENT WAS BEING TRANSFERRED TO THE TOILET WITH A CARE STAND AND 2 STAFF MEMBERS. THE PLASTIC CLASP ON THE BELT CAME APART WHILE THE RESIDENT WAS IN A STANDING POSITION. ACCORDING TO THE FACILITY THE RESIDENT FELL BACKWARDS AND HIT HER HEAD ON THE FLOOR. THIS RESULTED IN A HEAD LACERATION. THE RESIDENT WAS HOSPITALIZED WHERE A MRI REVEALED INTRACRANIAL BLEED. THE RESIDENT HAS SINCE BEEN RELEASED FROM THE HOSP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEDCARE STAND FNG MEDCARE PRODUCTS 400002

Patients

Seq Age Sex Outcome Treatment
1 74 YR Hospitalization