POSEY KEEPSAFE DELUXE
Report
- Report Number
- 2020362-2012-00067
- Event Type
- Malfunction
- Date Received
- February 10, 2012
- Report Date
- January 19, 2012
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). RESULTS: EVAL OF THE RETURNED PRODUCT FOUND THE ALARM IS MISSING THE HOLD/SUSPEND BUTTON. THERE IS BATTERY LEAKAGE ON THE BATTERY SPRINGS AND CORROSION ON THE BATTERY CONTACTS. THERE IS SIGNS OF WATER INTRUSION. THE UNIT WILL NOT POWER ON WITH BATTERIES OR 9V BATTERIES. NOTE: POSEY INSTRUCTIONS FOR USE HAS A STATEMENT: IF THE POSEY KEEPSAFE IS SUBJECTED TO SEVERE MECHANICAL SHOCK, SUCH AS DROPPING, OR IS SUBMERGED IN LIQUID, IT MAY STOP FUNCTIONING AND DESIGNED. VISUALLY INSPECT THE UNIT FOR CRACKS IN THE CASE, MISSING SCREWS, MISSING BATTERY DOOR, BROKEN WIRES OR SIGNS OF EXPOSURE TO LIQUID. AFTER EACH INCIDENT, YOU MUST VERIFY THAT THE UNIT IS WORKING PROPERLY. CHECK TO MAKE SURE THE AUDIBLE ALARM IS FUNCTIONING PROPERLY BY APPLYING AND LIFTING WEIGHT OFF SENSOR IN SEVERAL SPOTS TO ACTIVATE THE ALARM. DO NOT USE IF THE AUDIBLE ALARM DOES NOT SOUND. TAKE CARE WHEN INSTALLING NEW BATTERIES. THE ALARM WILL NOT WORK IF BATTERIES ARE INSTALLED IMPROPERLY. ALWAYS INSTALL A COMPLETELY NEW SET OF BATTERIES WHEN THE CHIRPING DUE TO LOW BATTERY POWER SOUND BEGINS. DO NOT REPLACE A SINGLE CELL, BUT ALL CELLS IN THE ALARM. (B)(4).
CUSTOMER REPORTED THAT WHILE THE ALARM WAS BEING TESTED TO BE USED ON A NEW PT, THE HOLD/SUSPEND BUTTON CAME OFF. THE CUSTOMER REPORTED THEY DID NOT MIS-HANDLE THE UNIT AND THE DAMAGED OCCURRED DURING NORMAL USE OF THE UNIT. NO OTHER DAMAGE REPORTED BY THE CUSTOMER. THERE WAS NO PT INCIDENT OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POSEY KEEPSAFE DELUXE | KMI | J. T. POSEY CO. | 8374 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | OVER-THE-MATTRESS SENSOR PAD: MODEL 8307, LOT# UNK |