FDA Adverse Event Death Summary report: N

POWER F2

MDR report key: 24689746 · Received March 25, 2026

Report

Report Number
1523574-2026-00029
Event Type
Death
Date Received
March 25, 2026
Report Date
March 25, 2026
Manufacturer
FERNO-WASHINGTON, INC.
Product Code
FPO
UDI-DI
00190790006570
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

THE END USER REPORTED AFTER ARRIVAL AT THE HOSPITAL WITH AN ACTIVE CARDIAC PATIENT; WHILE UNLOADING THE STRETCHER FROM THE AMBULANCE, THE LEGS OF THE STRETCHER HAD TO BE MANUALLY LOWERED BUT THEN THE CARRIAGE ARMS WOULD NOT LOWER DOWNWARD TO RELEASE THE STRETCHER FROM THE LOADER. THE CREW WAS ABLE TO MANUALLY LOWER THE CARRIAGE ARMS AND COMPLETE THE TRANSPORT INTO THE HOSPITAL. THE END USER ADVISED AN ESTIMATED TWO-MINUTE DELAY WAS EXPERIENCED DURING THE UNLOADING PROCESS. PATIENT DETAILS: MALE, APPROXIMATELY 220 LBS, PATIENT WAS EXPERIENCING AND BEING TREATED FOR CARDIAC ARREST ON SCENE AND THROUGHOUT TRANSPORT. CPR WAS BEING ADMINISTERED VIA LIFELINE ARM. THE PATIENT WAS LATER PRONOUNCED DECEASED AT THE HOSPITAL. THE END USER ADVISED THEY WERE UNABLE TO DUPLICATE THE ISSUE AFER THE INCIDENT AND MADE THE DECISION TO KEEP THE POWER F2 IN SERVICE, IN MANUAL MODE ONLY, UNTIL EVALUATION. THIS CONTINUES TO BE AN ACTIVE INVESTIGATION AND ADDITIONAL INFORMATION WILL BE PROVIDED IN A FOLLOW UP REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
532118 POWER F2 POWER F2 FPO FERNO-WASHINGTON, INC. 0567093 M255240 00190790006570

Patients

Seq Age Sex Outcome Treatment
1 NA Male Death