4D INTEGRATED TREATMENT CONSOLE
Report
- Report Number
- 2916710-2012-00008
- Date Received
- February 16, 2012
- Date of Event
- January 19, 2012
- Report Date
- January 19, 2012
- Manufacturer
- VARIAN MEDICAL SYSTEMS, INC
- Product Code
- IYE
- PMA / PMN Number
- K001643
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE PROBLEM WAS REPRODUCED. SAFETY ANALYSIS: THE TREATMENT FIELD IS NOT IMPACTED; TREATMENT FIELDS WORK WITHOUT ANY ISSUES. THIS IS AN ISSUE WITH THE AD-HOC IMAGE TEMPLATE. THE 4DITC AD-HOC CIAO (COMPLETE IRRADIATED AREA OUTLINE) CALCULATION RESULTS IN AN ERROR THAT WILL BE PRESENT IN THE 4DITC MLC DISPLAY. IF THE CIAO IMAGE GENERATED BY 4DITC IS PROJECTED ON THE PT USING THE LIGHT FIELD, THE INCORRECT FIELD EDGE WILL BE DISPLAYED. IF THE FIELD EDGE IS USED FOR SET-UP OF ANOTHER ABUTTING TREATMENT FIELD, THE AREA BETWEEN THE PROJECTED FIELD EDGE AND THE ABUTTING FIELD EDGE WOULD BE MISMATCHED AND NOT TREATED. THE MOST SERIOUS PROBABLE SCENARIO WOULD BE AN UNDER-DOSE OF THE PRESCRIBED AREA OF TREATMENT VOLUME. AN UNDER-DOSE CARRIED OUT FOR THE ENTIRE COURSE OF TREATMENT COULD LEAD TO LOSS OF TUMOR CONTROL AND POTENTIAL TUMOR SPREAD WHICH MIGHT LEAD TO REOCCURRENCE OR SPREAD. (B)(4). NO FURTHER F/U TO THIS MDR IS ANTICIPATED. (B)(4).
AT THE TIME OF FILM ACQUISITION, MLC SHAPE WAS DIFFERENT FROM PLAN. THE RADIATION TECHNOLOGIST OPENED A PLAN IN 4D INTEGRATED TREATMENT CONSOLE (4DITC), AND MODED UP THE FIELD OF THE FILM. THE MLC DISPLAY ON 4DITC WAS DIFFERENT FROM THE MLC SHAPE IN THE PLAN. WHEN THE FILM IS ADDED IN SCHEDULING OF RT CHART, THIS PROBLEM DOES NOT OCCUR. WHEN ADD FILM IN 4DITC IS USED, A PROBLEM OCCURS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 4D INTEGRATED TREATMENT CONSOLE | ACCELERATOR, LINEAR, MEDICAL | IYE | VARIAN MEDICAL SYSTEMS, INC | H51 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |