NONE
Report
- Report Number
- 2955842-2026-17670
- Event Type
- Malfunction
- Date Received
- March 25, 2026
- Date of Event
- March 2, 2026
- Report Date
- May 8, 2026
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- GCJ
- UDI-DI
- 00886874116555
- PMA / PMN Number
- K191736
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
INTUITIVE SURGICAL INC. (ISI) RECEIVED THE 30-DEGREE ENDOSCOPE PLUS FOR FAILURE ANALYSIS INVESTIGATION. THE ENDOSCOPE WAS ANALYZED AND WAS PLACED THROUGH FRICTION TEST USING A SCREENING AID TO MANUALLY ROTATE THE ADAPTER TO DETECT FOR FRICTION. THE CAMERA INSTRUMENT ADAPTER DID NOT ROTATE PROPERLY AND/OR NOISES WERE HEARD DURING TESTING AND CONFIRMED AS BINDING. ADDITIONAL OBSERVATIONS NOT REPORTED BY SITE: THE ENDOSCOPE CABLE INTEGRATED CONNECTOR WAS VISUALLY INSPECTED AND FOUND WITH THE PCA BOARD EXHIBITED MISSING ELECTRICAL CONTACTS. THE ENDOSCOPE WAS VISUALLY INSPECTED AND FOUND WITH COSMETIC DAMAGE. DURING INSPECTION OF THE ENDOSCOPE CABLE INTEGRATED CONNECTOR, THE CAB INTEGRATED CONNECTOR HOUSING EXHIBITED DISCOLORATION. VISUAL INSPECTION CONFIRMED. THE ENDOSCOPE WAS EVALUATED AND FOUND WITH A CAMERA INSTRUMENT ADAPTER DEFECT. THE ENDOSCOPE WAS PLACED THROUGH FRICTION TEST USING A SCREENING AIDE TO MANUALLY ROTATE THE ADAPTER TO DETECT FOR FRICTION. THE CAMERA INSTRUMENT ADAPTER DID NOT ROTATE PROPERLY AND/OR NOISES WERE HEARD DURING TESTING AND CONFIRMED AS BINDING. THE CAMERA INSTRUMENT ADAPTER REMOVED FROM ENDOSCOPE HOUSING AND EVALUATED AND FOUND WITH AEA SHAFT BEARING CONTRIBUTING TO FRICTION. THE ENDOSCOPE WAS VISUALLY INSPECTED AND WAS FOUND WITH DAMAGE AT THE DISTAL END. THE DISTAL WINDOW LOCATED AT DISTAL TIP WAS VISUALLY INSPECTED AND FOUND CRACKED. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS. THE ENDOSCOPE WAS PLACED THROUGH FRICTION TEST USING A SCREENING AID TO MANUALLY ROTATE THE ADAPTER TO DETECT FOR FRICTION. THE CAMERA INSTRUMENT ADAPTER DID NOT ROTATE PROPERLY AND/OR NOISES WERE HEARD DURING TESTING AND CONFIRMED AS BINDING. THE PROBABLE ROOT CAUSE OF THIS BINDING IS ATTRIBUTED TO COMPONENT SUSCEPTIBILITY TO INCREASED FRICTION.
INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE DA VINCI PRODUCT WITH AN ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/OR IF ADDITIONAL INFORMATION IS RECEIVED.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT AFTER A DA VINCI-ASSISTED PARASTOMAL HERNIA REPAIR SURGICAL PROCEDURE, THE 30-DEGREE ENDOSCOPE PLUS IMAGE WAS REVERSED OR UPSIDE DOWN. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 532149 | NONE | ENDOSCOPE PLUS | GCJ | INTUITIVE SURGICAL, INC | 470057-11 | N/A | 00886874116555 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES. |