FDA Adverse Event Malfunction Summary report: N

NONE

MDR report key: 24685677 · Received March 25, 2026

Report

Report Number
2955842-2026-17670
Event Type
Malfunction
Date Received
March 25, 2026
Date of Event
March 2, 2026
Report Date
May 8, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
GCJ
UDI-DI
00886874116555
PMA / PMN Number
K191736
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL INC. (ISI) RECEIVED THE 30-DEGREE ENDOSCOPE PLUS FOR FAILURE ANALYSIS INVESTIGATION. THE ENDOSCOPE WAS ANALYZED AND WAS PLACED THROUGH FRICTION TEST USING A SCREENING AID TO MANUALLY ROTATE THE ADAPTER TO DETECT FOR FRICTION. THE CAMERA INSTRUMENT ADAPTER DID NOT ROTATE PROPERLY AND/OR NOISES WERE HEARD DURING TESTING AND CONFIRMED AS BINDING. ADDITIONAL OBSERVATIONS NOT REPORTED BY SITE: THE ENDOSCOPE CABLE INTEGRATED CONNECTOR WAS VISUALLY INSPECTED AND FOUND WITH THE PCA BOARD EXHIBITED MISSING ELECTRICAL CONTACTS. THE ENDOSCOPE WAS VISUALLY INSPECTED AND FOUND WITH COSMETIC DAMAGE. DURING INSPECTION OF THE ENDOSCOPE CABLE INTEGRATED CONNECTOR, THE CAB INTEGRATED CONNECTOR HOUSING EXHIBITED DISCOLORATION. VISUAL INSPECTION CONFIRMED. THE ENDOSCOPE WAS EVALUATED AND FOUND WITH A CAMERA INSTRUMENT ADAPTER DEFECT. THE ENDOSCOPE WAS PLACED THROUGH FRICTION TEST USING A SCREENING AIDE TO MANUALLY ROTATE THE ADAPTER TO DETECT FOR FRICTION. THE CAMERA INSTRUMENT ADAPTER DID NOT ROTATE PROPERLY AND/OR NOISES WERE HEARD DURING TESTING AND CONFIRMED AS BINDING. THE CAMERA INSTRUMENT ADAPTER REMOVED FROM ENDOSCOPE HOUSING AND EVALUATED AND FOUND WITH AEA SHAFT BEARING CONTRIBUTING TO FRICTION. THE ENDOSCOPE WAS VISUALLY INSPECTED AND WAS FOUND WITH DAMAGE AT THE DISTAL END. THE DISTAL WINDOW LOCATED AT DISTAL TIP WAS VISUALLY INSPECTED AND FOUND CRACKED. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS. THE ENDOSCOPE WAS PLACED THROUGH FRICTION TEST USING A SCREENING AID TO MANUALLY ROTATE THE ADAPTER TO DETECT FOR FRICTION. THE CAMERA INSTRUMENT ADAPTER DID NOT ROTATE PROPERLY AND/OR NOISES WERE HEARD DURING TESTING AND CONFIRMED AS BINDING. THE PROBABLE ROOT CAUSE OF THIS BINDING IS ATTRIBUTED TO COMPONENT SUSCEPTIBILITY TO INCREASED FRICTION.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE DA VINCI PRODUCT WITH AN ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/OR IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT AFTER A DA VINCI-ASSISTED PARASTOMAL HERNIA REPAIR SURGICAL PROCEDURE, THE 30-DEGREE ENDOSCOPE PLUS IMAGE WAS REVERSED OR UPSIDE DOWN. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
532149 NONE ENDOSCOPE PLUS GCJ INTUITIVE SURGICAL, INC 470057-11 N/A 00886874116555

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES.