OT PING METER
Report
- Report Number
- 2939301-2012-01948
- Event Type
- Injury
- Date Received
- February 24, 2012
- Report Date
- February 7, 2012
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
THE LAY USER/PATIENT'S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE A LOW/DEAD BATTERY. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. THE 510(K) # IS K082590.
ON (B)(6) 2012, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING THAT HER ONETOUCH ULTRAPING METER DOES NOT TURN ON. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE CCA WAS ADVISED BY THE PATIENT THAT THE ALLEGED ISSUE BEGAN ON EITHER (B)(6) 2012 OR ON (B)(6) 2012 AT 2:00PM. THE PATIENT STATED THAT SHE MANAGES HER DIABETES WITH THE INSULIN PUMP AND "SKIPPED HER USUAL NOVOLOG INTAKE" IN RESPONSE TO THE REPORTED ISSUE. IMMEDIATELY AFTER THE ALLEGED ISSUE OCCURRED, THE PATIENT CLAIMED TO HAVE DEVELOPED SYMPTOMS OF "LIGHTHEADEDNESS, NAUSEA, AND SHAKINESS" BUT DENIED RECEIVING ANY TREATMENT IN RESPONSE TO THESE SYMPTOMS. DURING TROUBLESHOOTING, CCA NOTED THAT THE BATTERIES DID NOT NEED REPLACEMENT. REPLACEMENT METER WAS SENT TO THE PATIENT. ALTHOUGH THE PATIENT DENIED RECEIVING ANY MEDICAL TREATMENT AFTER THE ALLEGED ISSUE OCCURRED, THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT CLAIMED TO HAVE DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN INC. | 3232602 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | Life Threatening |