FDA Adverse Event Malfunction Summary report: N

FLEXIVA PULSE ID

MDR report key: 24678210 · Received March 24, 2026

Report

Report Number
2124215-2026-16199
Event Type
Malfunction
Date Received
March 24, 2026
Date of Event
February 27, 2026
Report Date
April 20, 2026
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
GEX
UDI-DI
08714729978688
PMA / PMN Number
K210925
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IS NOT AVAILABLE FOR ANALYSIS; THEREFORE, NO PHYSICAL ANALYSIS OF THE PRODUCT COULD BE PERFORMED. THE PRODUCT RECORD REVIEW (PRR) RESULTS DETAILED IN THE PRR TABLE DID NOT DETERMINE THE PROBABLE CAUSE OF THE COMPLAINT. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET SPECIFICATION PRIOR TO SHIPPING AND NO MANUFACTURING DEVIATIONS WERE NOTED THAT COULD HAVE CONTRIBUTED TO THE EVENT REPORTED. A RISK REVIEW WAS COMPLETED AND CONFIRMED THAT THE EVENT OF " FIBER BREAK" WAS DEFINED IN THE RISK DOCUMENTATION AND IS DOCUMENTED ACCORDINGLY IN THE PRR. THIS EVENT TYPE HAS BEEN ACCOUNTED FOR DURING PRODUCT RISK ANALYSIS TO SUPPORT ACCEPTABLE RISK BENEFIT FOR THE PRODUCT. A LABELING REVIEW WAS PERFORMED ON THE DEVICE INSTRUCTIONS FOR USE (IFU). THE IFU CITED THAT, THE FIBER SHOULD NOT BE CLAMPED WITH FORCEPS OR OTHER SECURING INSTRUMENTS AS IT COULD RESULT IN FIBER DAMAGE OR BREAKAGE. BASED ON THE INFORMATION AVAILABLE, THE CONCLUSION CODE "CAUSE NOT ESTABLISHED" HAS BEEN ASSIGNED AS THE MOST PROBABLE CAUSE FOR THE COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING URETEROSCOPY, THE FIBER BROKE AND STOPPED WORKING. AN ALTERNATE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING URETEROSCOPY, THE FIBER BROKE AND STOPPED WORKING. AN ALTERNATE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
275225 FLEXIVA PULSE ID POWERED LASER SURGICAL INSTRUMENT GEX BOSTON SCIENTIFIC CORPORATION M006L8406920 0037328858 08714729978688

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown