FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE

MDR report key: 24674862 · Received March 24, 2026

Report

Report Number
9610773-2026-02171
Event Type
Malfunction
Date Received
March 24, 2026
Date of Event
November 6, 2025
Report Date
March 24, 2026
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761074971
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE FOREIGN MATERIAL IN THE LUER-LOCK (LLK) PLUG OF THE VIDEO-CABLE SECTION IS CAUSE NOT ESTABLISHED AND THE MOST PROBABLE CAUSE OF THE DAMAGED FIBER BONDING IN THE OUTER TUBE AREA (DISTAL END) IS TRACED TO COMPONENT FAILURE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE GYNECOLOGIC LAPAROSCOPE EXHIBITED DAMAGED FIBER BONDING IN THE OUTER TUBE AREA (DISTAL END) AND FOREIGN MATERIAL IN THE LUER-LOCK (LLK) PLUG OF THE VIDEO-CABLE SECTION. NO PATIENTS WERE INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
737903 VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE LAPAROSCOPE, GYNECOLOGIC HET OLYMPUS WINTER & IBE GMBH WA50042A 04042761074971

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown