VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE
Report
- Report Number
- 9610773-2026-02159
- Event Type
- Malfunction
- Date Received
- March 24, 2026
- Date of Event
- December 8, 2025
- Report Date
- March 24, 2026
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- HET
- UDI-DI
- 04042761075015
- PMA / PMN Number
- K111788
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. BASED ON THE RESULTS OF THE INVESTIGATION, THE REPORTABLE EVENTS DEFORMED OUTER TUBE, THE CHARGED COUPLED DEVICE (R-UNIT) HAD A KINK WERE TRACED TO COMPONENT FAILURE. HOWEVER, A DEFINITIVE ROOT CAUSE FOR THE EVENT LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION CONTAINED FOREIGN MATERIAL COULD NOT BE ESTABLISHED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE SHOWED A DEFORMED OUTER TUBE, THE CHARGED COUPLED DEVICE (R-UNIT) HAD A KINK, PREVENTING DISASSEMBLY AND REASSEMBLY. ADDITIONALLY, THE LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION CONTAINED FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 737386 | VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE | RIGID VIDEO LAPAROSCOPE | HET | OLYMPUS WINTER & IBE GMBH | WA50052A | 04042761075015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |