FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE

MDR report key: 24674659 · Received March 24, 2026

Report

Report Number
9610773-2026-02159
Event Type
Malfunction
Date Received
March 24, 2026
Date of Event
December 8, 2025
Report Date
March 24, 2026
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761075015
PMA / PMN Number
K111788
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. BASED ON THE RESULTS OF THE INVESTIGATION, THE REPORTABLE EVENTS DEFORMED OUTER TUBE, THE CHARGED COUPLED DEVICE (R-UNIT) HAD A KINK WERE TRACED TO COMPONENT FAILURE. HOWEVER, A DEFINITIVE ROOT CAUSE FOR THE EVENT LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION CONTAINED FOREIGN MATERIAL COULD NOT BE ESTABLISHED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE SHOWED A DEFORMED OUTER TUBE, THE CHARGED COUPLED DEVICE (R-UNIT) HAD A KINK, PREVENTING DISASSEMBLY AND REASSEMBLY. ADDITIONALLY, THE LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION CONTAINED FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
737386 VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE RIGID VIDEO LAPAROSCOPE HET OLYMPUS WINTER & IBE GMBH WA50052A 04042761075015

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown