FDA Adverse Event Malfunction Summary report: N

BREATHING CIRCUITS

MDR report key: 24673952 · Received March 24, 2026

Report

Report Number
3001421318-2026-00157
Event Type
Malfunction
Date Received
March 24, 2026
Date of Event
February 26, 2026
Report Date
March 23, 2026
Manufacturer
HAMILTON MEDICAL AG
Product Code
CAI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG REF NR: COMP-(B)(4). INVESTIGATION OUTCOME: IT WAS REPORTED THAT WE HAVE HAD SEVERAL OCCURRENCES OF "EXHALATION OBSTRUCTED" ON OUR VENTS. THE DEVICE PASSES PRE-OP CHECKS BUT ONCE VENTILATION STARTS, IT ALARMS "EXHALATION OBSTRUCTED" . THERE WAS NO OBSTRUCTION TO THE EXPIRATORY BLEED PORT. WHEN TESTING THE DEVICE WITH THE CLINICAL ENGINEERING TEST EXPIRATORY VALVE, IT PASSES. TESTING WITH THE USER PROVIDED CIRCUITS (PRODUCT REF: 260128, LOT NO. 201868), THE TEST WOULD FAIL. I OPENED UP TWO MORE CIRCUITS BELONGING TO THE SAME LOT NUMBER AND THE RESULT WAS THE SAME. A BREATHING CIRCUIT BELONGING TO LOT NO. 201894 WAS ALSO TESTED AND IT PASSED. THE MOST PROBABLE ROOT CAUSE WAS A STICKY MEMBRANE OR FAULTY EXPIRATORY VALVE AS PART OF THE BREATHING CIRCUIT SET. HOWEVER, THIS COULD NOT BE CONFIRMED AS HAMILTON MEDICAL AG DID NOT RECEIVE THE BREATHING CIRCUITS ASSOCIATED WITH THIS SPECIFIC CASE. THE ISSUE WAS RESOLVED BY REPLACING THE AFFECTED BREATHING CIRCUIT SET WITH ANOTHER ONE. NO FURTHER INFORMATION WAS RECEIVED. THIS CASE IS CONSIDERED AS CLOSED.

Description of Event or Problem · 0

HAMILTON MEDICAL AG RECEIVED THE FOLLOWING EVENT DESCRIPTION: "WE HAVE HAD SEVERAL OCCURRENCES OF "EXHALATION OBSTRUCTED" ON OUR VENTS. THE DEVICE PASSES PRE-OP CHECKS BUT ONCE VENTILATION STARTS, IT ALARMS "EXHALATION OBSTRUCTED". THERE WAS NO OBSTRUCTION TO THE EXPIRATORY BLEED PORT. WHEN TESTING THE DEVICE WITH THE CLINICAL ENGINEERING TEST EXPIRATORY VALVE, IT PASSES. TESTING WITH THE USER PROVIDED CIRCUITS (PRODUCT REF: 260128, LOT NO. 201868), THE TEST WOULD FAIL. I OPENED UP TWO MORE CIRCUITS BELONGING TO THE SAME LOT NUMBER AND THE RESULT WAS THE SAME. A BREATHING CIRCUIT BELONGING TO LOT NO. 201894 WAS ALSO TESTED AND IT PASSED." NO HEALTH CONSEQUENCES OR IMPACT. CIRCUITS WERE CHANGED MULTIPLE TIMES TO CORRECT THE ISSUE. THE CASE HAS NOT BEEN REVIEWED YET BY HAMILTON MEDICAL AG. THEREFORE THE FAILURE DESCRIPTION COULD NOT BE CONFIRMED YET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
737695 BREATHING CIRCUITS BREATHING CIRCUITS CAI HAMILTON MEDICAL AG 260128

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown