FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 24664785 · Received March 23, 2026

Report

Report Number
9617229-2026-05111
Event Type
Injury
Date Received
March 23, 2026
Date of Event
February 24, 2026
Report Date
March 23, 2026
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
UDI-DI
10888628000254
PMA / PMN Number
P990074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS DEFLATION WAS RECEIVED ON APR 27, 2026, WITH LOT NUMBER 1760017. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION (IF APPLICABLE). PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: ¿ DEFLATION: OBSERVED AN OPENING THROUGH MICROSCOPIC INSPECTION ASSESSED AS FOLD FLAW OPENING. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (CREASES, WEAR ABRASION) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.

Additional Manufacturer Narrative · 0

FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. REASON FOR REOPERATION: DEFLATION.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
727736 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 1760017 10888628000254

Patients

Seq Age Sex Outcome Treatment
1 40 YR Female Required Intervention