FDA Adverse Event Malfunction Summary report: N

DA VINCI XI

MDR report key: 24658621 · Received March 22, 2026

Report

Report Number
2955842-2026-17191
Event Type
Malfunction
Date Received
March 22, 2026
Date of Event
February 25, 2026
Report Date
April 20, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874115374
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE ISSUE WAS RESOLVED BY CONTACTING THE TECHNICAL SUPPORT ENGINEER (TSE) AND FIELD SERVICE ENGINEER (FSE) WENT ON SITE AND CHANGED THE POWER SUPPLY. THE ALARM WAS ABLE TO BE RECOVERED AND THE SYSTEM WAS USABLE IN THIS STATE.

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE REPLACED THE MEDICAL GRADE POWER SUPPLY (MGPS) TO CORRECT THE REPORTED PROBLEM. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID NOT RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION, WHICH INDICATES THAT THE DEVICE MAY HAVE CONTRIBUTED TO THE CUSTOMER REPORTED ISSUE.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT AFTER A DA VINCI-ASSISTED SURGICAL PROCEDURE, AN INTUITIVE SURGICAL (IS) TECHNICAL SUPPORT ENGINEER (TSE) WAS CONTACTED FOR TROUBLESHOOTING ASSISTANCE TO REPORT THAT THE OPERATING ROOM (OR) ALARMED WHEN CONNECTING THE AC POWER CORD FROM THE CONSOLE. THE PROCEDURE WAS COMPLETED WITH NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY. THERE WERE NO DELAYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
425368 DA VINCI XI SURGEON SIDE CONSOLE, RECONDITIONED NAY INTUITIVE SURGICAL, INC 380723-12 N/A 00886874115374

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES.