FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 2465436 · Received February 16, 2012

Report

Report Number
1811755-2012-00516
Event Type
Malfunction
Date Received
February 16, 2012
Date of Event
January 24, 2012
Report Date
January 24, 2012
Manufacturer
STRYKER INSTRUMENTS, KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A MFR FIELD SERVICE TECH WAS DISPATCHED TO THE USER FACILITY TO EVALUATE THE DEVICE. VISUAL INSPECTION CONFIRMED THAT A WIRE FROM THE POWER CORD HAD DISCONNECTED FROM THE POWER PLUG ASSEMBLY. THE CAUSE OF THE DAMAGE IS UNK, BUT MAY BE THE RESULT OF MISHANDLING WHEN THE UNIT IS UNPLUGGED FROM THE WALL OUTLET. THE UNIT WAS REPAIRED AND RETURNED TO USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT THE END OF A SURGICAL PROCEDURE, AN EXPOSED WIRE FROM THE ROVER POWER CORD WAS OBSERVED NEAR THE POWER PLUG ASSEMBLY. NO PT OR USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE FYD STRYKER INSTRUMENTS, KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK