FDA Adverse Event
Malfunction
Summary report: N
NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE
MDR report key: 2465436
·
Received February 16, 2012
Report
- Report Number
- 1811755-2012-00516
- Event Type
- Malfunction
- Date Received
- February 16, 2012
- Date of Event
- January 24, 2012
- Report Date
- January 24, 2012
- Manufacturer
- STRYKER INSTRUMENTS, KALAMAZOO
- Product Code
- FYD
- PMA / PMN Number
- K012991
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
A MFR FIELD SERVICE TECH WAS DISPATCHED TO THE USER FACILITY TO EVALUATE THE DEVICE. VISUAL INSPECTION CONFIRMED THAT A WIRE FROM THE POWER CORD HAD DISCONNECTED FROM THE POWER PLUG ASSEMBLY. THE CAUSE OF THE DAMAGE IS UNK, BUT MAY BE THE RESULT OF MISHANDLING WHEN THE UNIT IS UNPLUGGED FROM THE WALL OUTLET. THE UNIT WAS REPAIRED AND RETURNED TO USE.
Description of Event or Problem · 1
IT WAS REPORTED THAT AT THE END OF A SURGICAL PROCEDURE, AN EXPOSED WIRE FROM THE ROVER POWER CORD WAS OBSERVED NEAR THE POWER PLUG ASSEMBLY. NO PT OR USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE | FYD | STRYKER INSTRUMENTS, KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |