SPACELABS ULTRAVIEW SL COMMAND MODULE
Report
- Report Number
- 3023361-2012-00006
- Event Type
- Malfunction
- Date Received
- February 23, 2012
- Date of Event
- January 28, 2012
- Report Date
- June 22, 2012
- Manufacturer
- SPACELABS MEDICAL INC.
- Product Code
- DSI
- PMA / PMN Number
- K050605
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE INITIAL TESTING CARRIED OUT BY THE SPACELABS FIELD SERVICE ENGINEER (FSE) AT THE CUSTOMER'S SITE CONFIRMED THE PATIENT MONITOR WAS OPERATING TO SPECIFICATIONS. SPACELABS IS WORKING WITH THE CUSTOMER TO COLLECT MORE DATA TO CONTINUE OUR INVESTIGATION. AT THIS TIME THERE IS INSUFFICIENT DATA TO IDENTIFY ROOT CAUSE OR DRAW ANY CONCLUSION. NO ONE HAS BEEN INJURED AS A RESULT OF THIS ALLEGED MALFUNCTION. WE WILL PROVIDE A SUPPLEMENTAL REPORT ONCE OUR INVESTIGATION IS COMPLETE.
TESTING CARRIED OUT BY THE SPACELABS FIELD SERVICE ENGINEER (FSE) ON SITE CONFIRMED THE DEVICE WORKED TO SPECIFICATIONS. A COPY OF WAVEFORM SHOWING AN EPISODE OF VENTRICULAR TACHYCARDIA WAS ANALYZED BY A SPACELABS ENGINEER. THE MORPHOLOGY OF THE WAVEFORM CHANGED FREQUENTLY AND SEVERAL LOW AMPLITUDE BEATS COULD HAVE CAUSED THE WAVEFORM TO NOT MEET THE CRITERIA FOR THE NUMBER OF CONSECUTIVE BEATS TO ALARM. WITH RESPECT TO THE REPORTED FAILURE TO ALARM FOR A LOW BLOOD PRESSURE CONDITION, AN ALARM WILL ONLY BE INITIATED WHEN BLOOD PRESSURE CONTINUOUSLY EXCEEDS THE ALARM THRESHOLD FOR A MINIMUM OF 3 SECONDS IN ORDER TO MINIMIZE NUISANCE ALARMS CAUSED BY TRANSIENT PRESSURE CHANGES. PRESSURE MEASUREMENTS ARE CALCULATED USING AN AVERAGING FUNCTION THAT IS BASED ON HEART RATE. IN THIS CASE, THE AVERAGE PRESSURE MEASUREMENT DID NOT EXCEED THE ALARM THRESHOLD FOR THE MINIMUM TIME REQUIRED FOR AN ALARM. NO ONE HAS BEEN INJURED AS A RESULT OF THIS ALLEGED MALFUNCTION. THE CUSTOMER IS CONTINUING TO USE THE DEVICE TO MONITOR PATIENTS. WE HAVE CONCLUDED THAT NO FURTHER ACTION IS REQUIRED AND CONSIDER THIS ISSUE CLOSED.
SPACELABS RECEIVED A REPORT THAT A PATIENT MONITOR FAILED TO ALARM ON A RUN OF VTAC AND LOW BLOOD PRESSURE IN THE INTENSIVE CARE UNIT.
SPACELABS RECEIVED A REPORT THAT A PATIENT MONITOR FAILED TO ALARM ON VTAC IN THE INTENSIVE CARE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPACELABS ULTRAVIEW SL COMMAND MODULE | ULTRAVIEW MULTIPARAMETER MODULE | DSI | SPACELABS MEDICAL INC. | 91496 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PATIENT MONITOR MODEL 91387, SN: (B)(4)| PATIENT MONITOR MODEL 91387, SN: (B)(4) |