FDA Adverse Event Malfunction Summary report: N

3.5X50MM BEVELED FT SCREW

MDR report key: 24650155 · Received March 20, 2026

Report

Report Number
1220246-2026-01568
Event Type
Malfunction
Date Received
March 20, 2026
Date of Event
February 26, 2026
Report Date
March 20, 2026
Manufacturer
ARTHREX, INC.
Product Code
HWC
UDI-DI
00888867360952
PMA / PMN Number
K210994
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.

Description of Event or Problem · 0

ON 02-MAR-2026, A FACILITY REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-8741-40 T10 BEVELED HEXALOBE DRIVER SHAFT BROKE OFF INTO THE HEAD OF THE SCREW WHEN INSERTING INTO PATIENT DURING A METACARPAL FX PROCEDURE. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT, AND THE CASE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS RECEIVED ON 04-MAR-2026. THE REP ADVISED THAT THE SURGEON USED A HANDLE (AR-8700RH) ATTACHED TO THE DRIVER SHAFT (AR-8741-40) TO INSERT THE SCREW BY HAND. THEY STATED THE SURGEON USED A 1.4MM GUIDE WIRE (AR-8741-14), AS WELL AS A 2.9MM DRILL BIT (AR-8741-25) TO PREP THE BONE FOR THE INSERTION OF THE SCREW. THE SURGEON WAS NOT ABLE TO USE THE DRIVER SHAFT (AR-8741-40) TO REMOVE THE SCREW SINCE THE DRIVER SHAFT BROKE OFF INTO THE HEAD OF THE SCREW (AR-8735BV-50). THEY WERE ABLE TO USE INSTRUMENTS THE HOSPITAL OWNED TO RETRIEVE THE REMAINING PIECES OF THE DRIVER SHAFT OUT OF THE HEAD OF THE SCREW. THEY THEN USED A DIFFERENT DRIVER SHAFT TO REMOVE THE SCREW (AR-8735BV-50), THEY RE-DRILLED WITH AN AR-8741-25, AND INSERTED AN AR-8735BV-48, WHICH WAS IMPLANTED USING A DIFFERENT DRIVER SHAFT. THE SURGEON USED THE 3.5MM BEVELED SCREWS, IMPLANTED/EXPLANTED AR-8735BV-50, AND IMPLANTED AR-8735BV-48.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714501 3.5X50MM BEVELED FT SCREW BONE FIXATION SCREW/PLATE HWC ARTHREX, INC. 3.5X50MM BEVELED FT SCREW UNK 00888867360952

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown