FDA Adverse Event
Malfunction
Summary report: N
HYPERFORM OCCLUSION BALLOON SYSTEM
MDR report key: 2464763
·
Received February 22, 2012
Report
- Report Number
- 2029214-2012-00075
- Event Type
- Malfunction
- Date Received
- February 22, 2012
- Date of Event
- January 27, 2012
- Report Date
- January 30, 2012
- Manufacturer
- EV3 NEUROVASCULAR
- Product Code
- NUF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE CATHETER AND GUIDEWIRE HAVE BEEN EVALUATED. THE EVALUATION SHOWED THAT THE GUIDEWIRE BROKE INTO TWO SEGMENTS (DUE TO TENSILE OVERLOAD) INSIDE THE CATHETER LUMEN AND ONE OF THE SEGMENTS PUNCTURED THE CATHETER LUMEN AT 1.5 CM FROM THE DISTAL TIP. CATHETER LUMEN. (B)(4).
Description of Event or Problem · 1
DURING BALLOON PREPARATION, IT WAS REPORTED THAT THE GUIDEWIRE WAS OBSERVED TO HAVE EXITED OUT OF THE CATHETER PRIOR TO THE DISTAL TIP. THE DEVICE WAS NOT USED IN THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HYPERFORM OCCLUSION BALLOON SYSTEM | OCCLUSION BALLOON CATHETER | NUF | EV3 NEUROVASCULAR | 104-4470 | 9495839 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |