FDA Adverse Event Injury Summary report: N

DORO® SKULL CLAMP

MDR report key: 24647294 · Received March 20, 2026

Report

Report Number
3003923584-2026-00008
Event Type
Injury
Date Received
March 20, 2026
Date of Event
February 9, 2026
Report Date
March 20, 2026
Manufacturer
BLACK FOREST MEDICAL GMBH
Product Code
HBL
UDI-DI
04250435501689
PMA / PMN Number
K001808
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE REPORTING CUSTOMER PROVIDED VIDEO FOOTAGE, WHICH, UPON REVIEW, SHOWED THAT THE TWO-PIN-HOLDER ADAPTER PLATE EXHIBITED ROTATIONAL MOVEMENT. THE ROCKER ARM APPEARS RIGID IN THE VIDEO AND SHOWS NO ROTATIONAL MOVEMENT, WHICH INDICATES PROPER FIXATION OF THE OPENLOCK MECHANISM. UPON RECEIPT OF THE DEVICE IN QUESTION AND THE ASSOCIATED ACCESSORIES (TWO-PIN-HOLDER ADAPTER PLATE), THE DEVICE WAS TESTED IN THE CONFIGURATION SHOWN IN THE VIDEO PROVIDED (REPORTEDLY USED WHEN THE LACERATION OCCURRED). ROTATIONAL MOVEMENT OF TWO-PIN-HOLDER ADAPTER PLATE CORRESPONDING TO THE PROVIDED VIDEO FOOTAGE WAS FOUND TO BE ONLY POSSIBLE WHEN THE RECEPTACLE (CLAMPING MECHANISM) FOR ATTACHMENT OF THE TWO-PIN-HOLDER ADAPTER PLATE IS NOT LOCKED OR INADEQUATELY LOCKED. NO DEVIATIONS WERE FOUND ON THE DEVICE. AT THE TIME OF THIS REPORT, THE CUSTOMER HAS BEEN INFORMED OF THE INVESTIGATION RESULTS AND THE CORRESPONDING INSTRUCTIONS REGARDING PROPER INSTALLATION OF THE DEVICE AND SAFETY CHECKS PRIOR TO CLINICAL USE AS PART OF THE COMPLAINT CORRESPONDENCE. FURTHER INFORMATION HAS BEEN REQUESTED. ANY NEW FINDINGS REGARDING THIS INCIDENT MAY BE INCLUDED IN A FOLLOW-UP REPORT UPON RECEIPT.

Description of Event or Problem · 0

CUSTOMER INFORMED US THAT ONE OF OUR PRODUCTS WAS INVOLVED IN A CASE IN WHICH THE PATIENT SUSTAINED A LACERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714568 DORO® SKULL CLAMP DORO® QR3 MULTI PURPOSE SKULL CLAMP HBL BLACK FOREST MEDICAL GMBH 3020-50 04250435501689

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other