TRILOGY EV300, USA
Report
- Report Number
- 2518422-2026-009795
- Event Type
- Malfunction
- Date Received
- March 19, 2026
- Date of Event
- March 12, 2026
- Report Date
- May 28, 2026
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- CBK
- UDI-DI
- 00606959052017
- PMA / PMN Number
- K181166
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE MANUFACTURER PREVIOUSLY RECEIVED INFORMATION ALLEGING THAT A TRILOGY EV300 DEVICE FAILED THE ACTIVE EXHALATION TEST PRIOR TO IMPLEMENTATION OF A FIELD CORRECTIVE ACTION (FCA) OR FIELD CHANGE ORDER (FCO). THERE WAS NO REPORT OF PATIENT HARM. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT. DURING EVALUATION AT THE MANUFACTURER'S SERVICE CENTER, THE REPORTED ISSUE WAS CONFIRMED. THE SERVICE TECHNICIAN REPLACED THE PROPORTIONAL VALVE (AECM) AND 3-WAY SOLENOID VALVE IN ACCORDANCE WITH THE FIELD SERVICE MANUAL. FOLLOWING REPAIR, THE DEVICE PASSED ALL REQUIRED PERFORMANCE AND VERIFICATION TESTING AND WAS RETURNED TO SERVICE. THE REMOVED PROPORTIONAL VALVE (AECM) AND 3-WAY SOLENOID VALVE WERE RETURNED TO THE MANUFACTURER'S PRODUCT INVESTIGATION LABORATORY (PIL) FOR FURTHER EVALUATION. THE PROPORTIONAL VALVE WAS INSTALLED ON A TRILOGY EVO STB UNIT AND TESTED USING THE TRILOGY EVO MULTIFUNCTIONAL TEST STATION. THE COMPONENT PASSED AECM VERIFICATION TESTING AT PRETEST AND POSTTEST. THE INVESTIGATION CONCLUDED THAT THE PROPORTIONAL VALVE OPERATED AS DESIGNED. THE 3-WAY SOLENOID VALVE WAS ALSO INSTALLED ON A TRILOGY EVO STB UNIT AND EVALUATED USING THE TRILOGY EVO MULTIFUNCTIONAL TEST STATION. THE COMPONENT PASSED AECM VERIFICATION AND AECM SOLENOID CURRENT VERIFICATION TESTING AT POSTTEST; HOWEVER, IT FAILED AECM VERIFICATION TESTING AT PRETEST. THE RESULTS ARE INDICATIVE OF A FAULTY SOLENOID VALVE. THE INVESTIGATION BY ENGINEERING IS ONGOING.
THE MANUFACTURE RECEIVED INFORMATION THAT AN EV300 USA DEVICE WAS BEING USED TO CREATE A TREATMENT PLAN FOR RESPIRATORY ASSIST AND NEEDS O2 CALIBRATION. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE WAS NOT REPORTED TO BE IN PATIENT USE. DURING THE EVALUATION OF THE TRILOGY EV300, USA DEVICE BY THE FIELD SERVICE ENGINEER (FSE), THE DEVICE FAILED THE SENSOR DRIFT VERIFICATION. THE FSE REPORTED THE FAILURE OF THE FLOW SENSOR, ACTIVE EXHALATION CONTROL MODULE (AECM) AND THE 3-WAY VALVE RESULTED IN INACCURATE READINGS, WHICH CAUSED THE DEVICE TO FAIL TESTING. REPLACING THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE, 3-WAY SOLENOID VALVES AND FLOW SENSOR RESOLVED THE ISSUE. A FINAL REPORT IS BEING SUBMITTED. IF NEW INFORMATION BECOMES AVAILABLE FOLLOWING THE COMPLETION OF THE DEVICE REPAIR THAT CHANGES THE OUTCOME OF THE INVESTIGATION AND ASSOCIATED MEDICAL DEVICE REPORTING A FOLLOW UP/SUPPLEMENTAL REPORT WILL BE COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 331179 | TRILOGY EV300, USA | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS, INC. | DS2200X11B | 00606959052017 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |