UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM
Report
- Report Number
- 1061932-2012-00487
- Event Type
- Malfunction
- Date Received
- February 22, 2012
- Date of Event
- January 23, 2012
- Report Date
- January 25, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K081930
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
SERVICE WAS NOT DISPATCHED FOR THIS EVENT. CONTROLS WERE RUN BEFORE AND AFTER THE INCIDENT AND THE INSTRUMENT IS CURRENTLY PERFORMING WITHIN QC SPECIFICATIONS. THE APPLICATION SPECIALIST RAN S-CAL TO CHECK INSTRUMENT CALIBRATION. THE REFERENCE VALUE S-CAL WAS 12.6 AND S-CAL CHECK RESULT WAS 12.7. THE CUSTOMER AT PRESENT DOES NOT SEND R FLAGS OR H&H MESSAGES FROM REMISOL TO THEIR LIS, AND THEREFORE MISSED THE H&H FAILURE. THE CUSTOMER IS LOOKING AT HOW THIS MESSAGE CAN BE RECEIVED BY THERE LIS TO PREVENT PROBLEMS IN THE FUTURE. THE ROOT CAUSE IS UNKNOWN, BUT MAY BE SAMPLE RELATED.
A CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT ERRONEOUSLY LOW HEMOGLOBIN (HGB) RESULTS OF 11.9 AND 11.8 G/DL WERE GENERATED BY UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM FOR TWO SPECIMENS FROM A KNOWN POLYCYTHAEMIC PATIENT. AN OPERATOR DEFINED H&H CHECK FAILED ERROR MESSAGE WAS OBSERVED BUT THERE WAS NO INSTRUMENT GENERATED MESSAGES FOR HGB PARAMETER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. TESTING ON AN ALTERNATE INSTRUMENT (COULTER LH500 HEMATOLOGY ANALYZER) PRODUCED HIGHER HGB RESULTS OF 12.5 AND 12.4 G/DL, WHICH WERE CONFIRMED BY A REFERENCE METHOD (HEMOCUE). HGB RESULT OF 12.5 G/DL WAS REPORTED OUT OF THE LABORATORY. ADDITIONALLY, A DISCREPANCY IN PLATELET COUNTS WAS NOTED BETWEEN THE DXH800 AND LH500 INSTRUMENTS, BUT THE DXH800 RESULTS OF 717 AND 689 (X 10^3 CELLS/UL) WERE CONFIRMED BY THE REFERENCE FLOW CYTOMETRY RESULTS OF 672 AND 721(X 10^3 CELLS/UL) AND WERE CONSIDERED CORRECT. LH500 FLAGGED THE PLATELET RESULTS AND THE RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT DUE TO THE ERRONEOUS RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER, INC. | DXH 800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR |