MASS SPECTROMETER, CLINICAL USE
Report
- Report Number
- 3009715051-2026-00001
- Event Type
- Malfunction
- Date Received
- March 19, 2026
- Date of Event
- February 20, 2026
- Report Date
- March 18, 2026
- Manufacturer
- AB SCIEX PTE. LTD.
- Product Code
- DOP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE COMPLAINT WAS INITIALLY REPORTED AS "OVERHEATED SOURCE AND OPEN FLAME OBSERVED". AN INITIAL INVESTIGATION WAS INITIATED BASED ON CUSTOMER COMMUNICATION/INTERVIEWS: SIMILAR SAMPLES WERE RUN PREVIOUSLY ON THIS INSTRUMENT WITH NO REPORTED INCIDENT. BEFORE THIS PARTICULAR RUN, THE FITTING WAS CHECKED AND THERE WAS NO LOOSENESS OF THE CAPILLARY OBSERVED. NO SOLVENT LEAKS WERE NOTICED DURING THE APPLICATION OF THE SAMPLE. WHEN THE TECHNICIAN NOTICED THE FLAME, THE PEEK TUBING WAS DETACHED AT THE ION SOURCE INTERFACE AND SOLVENT WAS DRIPPING ONTO THE FLOOR AWAY FROM THE ION SOURCE. THE TECHNICIAN PUT THE INSTRUMENT INTO STANDBY MODE, AT WHICH POINT THE FLAME STOPPED. FURTHER MANUFACTURER INVESTIGATION IS PENDING RECEIPT OF THE MALFUNCTIONED COMPONENT.
COMPLAINT OPENED TO ADDRESS FLAME OBSERVED ON THE [ION] SOURCE OF THE MASS SPECTROMETER. A FIELD SERVICE EMPLOYEE EVALUATED THE INSTRUMENT AND CONFIRMED THAT THE INCIDENT WAS LIMITED TO THE ION SOURCE ASSEMBLY. BURN MARKS (DISCOLOURATION) AND SLIGHT DEFORMATION TO THE TOP OF THE PROBE OF THE ION SOURCE ASSEMBLY WERE OBSERVED. NO SUBSTANTIAL DAMAGE OCCURRED TO THE REST OF THE INSTRUMENT OR TO THE CUSTOMER ENVIRONMENT. THE [ELECTRODE ADJUSTMENT] NUT WAS REPLACED, AND SUBSEQUENT TESTING WITH THE REPLACEMENT WAS PERFORMED AND CONFIRMED THAT THE ISSUE WAS RESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 704429 | MASS SPECTROMETER, CLINICAL USE | DOP | AB SCIEX PTE. LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |