WOUND DRAIN
Report
- Report Number
- 1018233-2012-00109
- Event Type
- Injury
- Date Received
- February 17, 2012
- Date of Event
- December 18, 2011
- Report Date
- January 19, 2012
- Manufacturer
- PRODUCTOS PARA EL CUIDADO DE LA SALUD
- Product Code
- GBX
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
THE RETURNED SAMPLE WAS FOUND WITH JAGGED AND SHARP EDGES. STRESS MARKS WERE NOTED AT THE BREAKAGE SITE WHICH IS INDICATIVE OF EXCESSIVE FORCE HAVING BEEN APPLIED TO THE DRAIN BEYOND ITS TENSILE CAPABILITIES. THE INNER DIAMETER AND THE OUTER DIAMETER WERE MEASURED AND FOUND WITHIN SPECIFICATION. THE LOT NUMBER IS UNKNOWN; THEREFORE, THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. THE INSTRUCTIONS FOR USE STATES THE FOLLOWING WARNINGS: "TO AVOID THE POSSIBILITY OF DRAIN DAMAGE OR BREAKAGE: AVOID SUTURING THROUGH DRAINS. DRAINS SHOULD LIE FLAT AND IN LINE WITH THE SKIN EXIT AREAS. PARTICULAR CARE SHOULD BE TAKEN TO AVOID ANY OBSTACLES TO THE DRAIN EXIT PATH. DRAINS SHOULD BE CHECKED FOR FREE MOTION DURING CLOSURE TO MINIMIZE THE POSSIBILITY OF BREAKAGE. DRAIN REMOVAL SHOULD BE DONE GENTLY BY HAND. DRAINS SHOULD NOT BE HANDLED WITH POINTED, TOOTHED OR SHARP INSTRUMENTS WHICH WOULD CAUSE CUTS OR NICKS AND LEAD TO SUBSEQUENT STRUCTURAL FAILURE OF THE DRAIN. SURGICAL REMOVAL MAY BE NECESSARY IF DRAIN IS DIFFICULT TO REMOVE OR BREAKS." (B)(4).
IT WAS REPORTED THAT THE WOUND DRAINAGE TUBE WAS FOUND BROKEN. THE TUBE WAS PLACED AS OF (B)(6) 2011 FOR AN OPERATION AND IT WAS REMOVED ON (B)(6) 2011. NO RESISTANCE WAS NOTICED WHEN THE REMOVAL. DURING AN X-RAY, IT WAS FOUND THAT THE BROKEN PART OF THE DRAIN TUBE REMAINED IN THE PATIENT'S BODY. THE DRAIN TUBE WAS REMOVED RIGHT AWAY BY ANOTHER OPERATION. THE PATIENT IS RECOVERING AND NO ISSUES WERE NOTED WITH THE SECOND REMOVAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WOUND DRAIN | GBX | PRODUCTOS PARA EL CUIDADO DE LA SALUD | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention |