INTERSTIM¿
Report
- Report Number
- 3004209178-2026-04844
- Event Type
- Malfunction
- Date Received
- March 18, 2026
- Date of Event
- January 5, 2024
- Report Date
- March 18, 2026
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- EZW
- UDI-DI
- 00613994913654
- PMA / PMN Number
- P970004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR URINARY DYSFUNCTION/SACRAL NERVE STIM AND GASTROINTESTINAL/ PELVIC FLOOR. IT WAS REPORTED THAT THEY WERE NEEDING BRAIN MRI. THE HEALTHCARE PROVIDER (HCP) THAT THEY ASSUME THAT THE PATIENT (PT) WAS REFERRING TO THE INS NOT WORKING. CALLER THEN SPOKE WITH PT'S NURSE WHO STATED THAT THE PT HAD AN MRI A WHILE BACK AND THE INS WAS PUT INTO MRI MODE BEFOREHAND HOWEVER ONCE THEY PT GOT BACK HOME THEY COULDN'T GET THE INS OUT OF MRI MODE. HCP SUSPECTS THIS HAPPENED ON (B)(6) 2024 WHEN PT HAD A CERVICAL MRI DONE. AGENT SUGGESTED TROUBLESHOOTING WITH EXTERNAL EQUIPMENT, REVIEWED MRI COMPATIBILITY INFO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 695052 | INTERSTIM¿ | STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON | EZW | MEDTRONIC PUERTO RICO OPERATIONS CO. | 3058 | 00613994913654 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Male |