FDA Adverse Event Malfunction Summary report: N

DIMENSION® CLINICAL CHEMISTRY SYSTEM

MDR report key: 2463251 · Received February 21, 2012

Report

Report Number
2517506-2012-00031
Event Type
Malfunction
Date Received
February 21, 2012
Date of Event
January 31, 2012
Report Date
February 1, 2012
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
CGZ
PMA / PMN Number
K970330
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED CHLORIDE RESULT WAS SAMPLE RELATED. THE IFU FOR THE DIMENSION QUIKLYTE INTEGRATED MULTISENSOR CONTAINS THE FOLLOWING INFORMATION: "FOLLOW THE INSTRUCTIONS WITH YOUR SPECIMEN COLLECTION DEVICE FOR COLLECTION, TUBE HANDLING, SPIN TIMES, AND G-FORCE ESPECIALLY WHEN USING PLASTIC BLOOD COLLECTION TUBES. FAILURE TO FOLLOW THE BLOOD COLLECTION TUBE INSTRUCTIONS MAY RESULT IN INCREASED MAINTENANCE OR TROUBLESHOOTING OF THE IMT SYSTEM." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED CHLORIDE RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE ORIGINAL SAMPLE WAS REPEATED AND A LOWER RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED CHLORIDE RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION® CLINICAL CHEMISTRY SYSTEM QUIKLYTE¿ INTEGRATED MULTISENSOR CGZ SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW

Patients

Seq Age Sex Outcome Treatment
1