FDA Adverse Event Malfunction Summary report: N

ESOPHYX2 - LINK DESIGN - SHORT

MDR report key: 2463142 · Received February 21, 2012

Report

Report Number
3005473391-2012-00053
Event Type
Malfunction
Date Received
February 21, 2012
Date of Event
January 23, 2012
Report Date
January 23, 2012
Manufacturer
ENDOGASTRIC SOLUTIONS, INC
Product Code
ODE
PMA / PMN Number
K092400
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FINDINGS: BASED ON THE PRODUCT AND ENGINEERING INVESTIGATION, NO MATERIAL OR PROCESS DEFECT WAS FOUND. THE CENTER PIN ON THE HELIX BROKE BECAUSE OF A DUCTILE FAILURE CAUSED BY TORSIONAL OVERLOAD. DURING RETRACTOR RETRIEVAL, A SIGNIFICANT TORSIONAL LOAD WAS APPLIED TO THE CENTERING NEEDLE FROM TWISTING AND DISTORTING THE HELIX. THERE WAS NO LATENT DEFECT PRECIPITATING THE FAILURE. ALTHOUGH THERE WAS NO INDICATION OF AN INJURY TO THE PATIENT OR A MANUFACTURING PROCESS OR MATERIALS ISSUE, THIS FAILURE MODE IS BEING REPORTED DUE TO ESTABLISHING IF THIS MALFUNCTION WERE TO REOCCUR, A SERIOUS INJURY MAY OCCUR.

Description of Event or Problem · 1

TOWARDS THE END OF A SUCCESSFUL TRANSORAL INCISIONLESS FUNDOPLICATION (TIF) PROCEDURE, THE CENTER PIN OF THE HELIX WAS BENT OUTSIDE THE HELICAL COIL. BECAUSE OF THIS, THE SURGEON EXPERIENCED DIFFICULTY RETRACTING THE HELIX BACK INTO THE TISSUE MOLD. AFTER MANY FAILED FORCEFUL ATTEMPTS TO RETRACT THE HELIX, THE CENTER PIN FINALLY BROKE AND ADHERED TO THE SURFACE OF THE TISSUE MOLD. IN AN ATTEMPT TO RETRIEVE THE PIN WITH A GRASPER, THE PIN WAS INADVERTENTLY BUMPED AND THE PIN DROPPED INTO THE STOMACH. SUCTION WAS USED TO RETRIEVE THE PIN, THOUGH VISUAL CONFIRMATION OF THE RECOVERY COULD NOT BE VERIFIED. AS A PRECAUTION, AN X-RAY WAS TAKEN OF THE GENERAL AREA WITH NO EVIDENCE THE CENTER PIN REMAINED IN THE PATIENT, SO THE SURGEON'S ASSESSMENT WAS THE PIN WAS SUCCESSFULLY SUCTIONED OUT. THERE WERE NO PATIENT ADVERSE EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESOPHYX2 - LINK DESIGN - SHORT ODE ODE ENDOGASTRIC SOLUTIONS, INC R2002 401159

Patients

Seq Age Sex Outcome Treatment
1 72 YR