PRIM SYM SET 2 CL W/FILTERTEK VALVE
Report
- Report Number
- 9615050-2012-00123
- Event Type
- Malfunction
- Date Received
- February 14, 2012
- Date of Event
- January 1, 2012
- Report Date
- January 18, 2012
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FPA
- PMA / PMN Number
- UNK
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THE LIST NUMBER HAS A COMMON DEVICE NAME OF 80FPA AND A 510K OF K103344. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).
THE CUSTOMER CONTACT REPORTED BACKFLOW OF FLUID FROM THE SECONDARY SOLUTION CONTAINER INTO THE PRIMARY SOLUTION CONTAINER. THE PRIMARY TUBING SET WAS BEING USED TO DELIVERY AN UNSPECIFIED VOLUME OF NORMAL SALINE, AT AN UNSPECIFIED RATE, VIA A SYMBIQ PUMP. AT AN UNSPECIFIED TIME, THE MALE ADAPTER OF A SECONDARY TUBING SET WAS CONNECTED TO THE PROXIMAL CLAVE Y-SITE OF THE PRIMARY TUBING SET FOR PIGGYBACK DELIVERY OF 100ML OF AN UNSPECIFIED CONCENTRATION OF CEPHAZOLIN. THE PRIMARY SOLUTION CONTAINER WAS LOWERED BELOW THE SECONDARY SOLUTION CONTAINER. IT WAS REPORTED THAT 10 MINS AFTER THE DELIVERY WAS STARTED, THE NURSE NOTED THAT SECONDARY SOLUTION STARTED, THE NURSE NOTED THAT THE SECONDARY SOLUTION CONTAINER WAS EMPTY. THE PUMP AND TUBING SET WERE REPLACED AND THE THERAPY WAS RESUMED. DURING TESTING AT THE USER FACILITY, BACKFLOW OF SOLUTION FROM THE SECONDARY SOLUTION CONTAINER INTO THE PRIMARY SOLUTION CONTAINER WAS NOTED. THERE WERE NO REPORTED ADVERSE PT EFFECTS AND NO REPORTED DELAY OF THERAPY CRITICAL TO THIS PT. NO MEDICAL INTERVENTIONS WERE REQUIRED. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRIM SYM SET 2 CL W/FILTERTEK VALVE | UNK | FPA | HOSPIRA COSTA RICA LTD. | NA | UNK5H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | LOT #UNK| UNSPECIFIED SECONDARY TUBING SET: LIST #UNK, |