I-STAT TROPONIN (HS-TNI ) CARTRIDGE
Report
- Report Number
- 2245578-2026-00058
- Event Type
- Malfunction
- Date Received
- March 18, 2026
- Date of Event
- March 5, 2026
- Report Date
- April 24, 2026
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- PMA / PMN Number
- K240984
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 22-APR-2026. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AP (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR HS-TNI CARTRIDGE LOT B26016B.
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
NA.
ON (B)(6) 2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULTS ON A 40-YEAR-OLD FEMALE PATIENT WITH ABDOMINAL PAIN AND DIARRHEA. THERE WAS NO ADDITIONAL PATIENT INFORMATION. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: DATE : COLLECTED TESTED: RESULT : SAMPLE POS/NEG: I-STAT, (B)(6) 0026 , 1443 , 1507 , >1000 NG/L , A, WB POS, I-STAT , (B)(6) 0026 , 1539 , 1557 , >1000 NG/L, A, WB POS, ATELLIC, (B)(6) 0026, 1912 , 1939 , 3 PG/ML, A WB NEG. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. 99TH PERCENTILE 90% CI SEX N (NG/L, PG/ML) (NG/L, PG/ML), FEMALE 490 13 (10, 17), MALE 404 28 (19, 58), OVERALL 895 21 (14, 30).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 370718 | I-STAT TROPONIN (HS-TNI ) CARTRIDGE | HS-TNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | B26016B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |