FDA Adverse Event
Malfunction
Summary report: N
CERAMO PLANO-S MICRO SCISSORS
MDR report key: 24625294
·
Received March 18, 2026
Report
- Report Number
- 3001286223-2026-00002
- Event Type
- Malfunction
- Date Received
- March 18, 2026
- Report Date
- March 18, 2026
- Manufacturer
- FEHLING INSTRUMENTS GMBH
- Product Code
- DWS
- UDI-DI
- 04058749009477
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
THE INVESTIGATION HAS REVEALED THAT THE BREAKAGE WAS LIKELY CAUSED BY OVERLOADING DUE TO EXCESSIVE OPERATING FORCE. NO HARM WAS CAUSED TO THE PATIENT. THE INSTRUMENT IS BEYOND REPAIR. NO FURTHER ACTION IS REQUIRED AT THIS TIME.
Description of Event or Problem · 0
A COMPLAINT WAS FILED REGARDING A PAIR OF MICRO SCISSORS; ACCORDING TO THE CUSTOMER, THE TIP HAD BROKEN OFF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 688301 | CERAMO PLANO-S MICRO SCISSORS | MICRO SCISSORS | DWS | FEHLING INSTRUMENTS GMBH | BSH-6 | APA | 04058749009477 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |