FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 2462078 · Received February 14, 2012

Report

Report Number
2245578-2012-00050
Event Type
Malfunction
Date Received
February 14, 2012
Date of Event
August 23, 2011
Report Date
February 2, 2012
Manufacturer
ABBOTT POINT OF CARE, INC.
Product Code
MMI
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN INVESTIGATION WAS COMPLETED (B)(4) 2012. RETAIN TESTING FOR LOTS T11118 AND T11124 WERE PART OF A LARGER STUDY THAT WAS INITIATED TO BETTER UNDERSTAND AND DETERMINE THE EFFECT OF ALTITUDE ON QUALITY CHECK CODES (ANALYZER DETECTED ERRORS) ON T LOTS OF TROPONIN. CARTRIDGE LOTS TESTED AT 5000 AND 10000 FEET ABOVE SEA LEVEL SHOWED INCREASE RATES OF ANALYZER DETECTED ERRORS.

Description of Event or Problem · 1

ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT ARE YIELDING A HIGH RATE FOR QUALITY CHECK CODES (QCC) 123, 126, AND 142 ON A PATIENT. THE CUSTOMER SUSPECTS THAT ALTITUDE MAY BE THE CAUSE. THE CUSTOMER STATS THAT THE EQUIPMENT IS PERFORMING AS INTEND WITH OTHER TYPES OF CARTRIDGES. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT. LOT NUMBER: T11118, MANUFACTURED: (B)(6) 2011, EXPIRES: 11/14/2011. LOT NUMBER: T11124, MANUFACTURED: (B)(6) 2011, EXPIRES: 11/28/2011. NOTE: ALL "T" LOTS ASSOCIATED WITH THIS INCIDENT HAVE EXPIRED ANY PREVIOUSLY ASSOCIATED WITH AN UNRELATED CORRECTIVE ACTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE, INC. NA T1118

Patients

Seq Age Sex Outcome Treatment
1 UNK