FDA Adverse Event Injury Summary report: N

IMP,TSV,4.7,8,MTX,MG

MDR report key: 24616019 · Received March 17, 2026

Report

Report Number
0001038806-2026-01434
Event Type
Injury
Date Received
March 17, 2026
Date of Event
February 25, 2026
Report Date
April 14, 2026
Manufacturer
ZIMVIE US CORP LLC
Product Code
DZE
UDI-DI
00889024020030
PMA / PMN Number
K101977
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN/NOT PROVIDED. D6A: IMPLANT DATE UNKNOWN/NOT PROVIDED. G4: ADDITIONAL PMA/510(K) NUMBER K101977.

Additional Manufacturer Narrative · 0

ZIMVIE DID NOT RECEIVE ONE (1) TSVTWB8, (IMP,TSV,4.7,8,MTX,MG) FOR EVALUATION. VISUAL INSPECTION COULD NOT BE PERFORMED. THE INVESTIGATION HAS BEEN PERFORMED BASED ON THE AVAILABLE INFORMATION USING THE ITEM NUMBER, DEVICE HISTORY RECORD (DHR) REVIEW, AND RISK MANAGEMENT FILE (RMF). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1285666. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1285666 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION WAS NOT ESTABLISHED. WITHOUT DEVICE RECEIPT, THE REPORTED EVENT IS NON-VERIFIABLE. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT G3: DATE RECEIVED BY MANUFACTURER. G6: CHECKED "FOLLOW-UP". H6: ENTERED EVALUATION CODES. H11: ADDED MANUFACTURER NARRATIVE.

Additional Manufacturer Narrative · 0

ZIMVIE RECEIVED ONE (1) TSVTWB8, (IMP,TSV,4.7,8,MTX,MG) FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED; THE IMPLANT SHOWS EXTENSIVE SIGNS OF WEAR/USE. SOME DAMAGE IS LIKELY CAUSED DURING REMOVAL. BONE DEBRIS OBSERVED ON ITS EXTERNAL THREADS. FRACTURE IDENTIFIED AT THE COLLAR. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1285666. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1285666 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE IMPLANT WAS FRACTURED AT THE COLLAR. THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #19 WAS REMOVED AS IT WAS FRACTURED. PT PRESENTED WITH FRACTURED IMPLANT.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283636 IMP,TSV,4.7,8,MTX,MG DENTAL IMPLANT DZE ZIMVIE US CORP LLC 1285666 00889024020030

Patients

Seq Age Sex Outcome Treatment
1