FDA Adverse Event Death Summary report: N

MONITOR DOCKING STATION

MDR report key: 2460559 · Received February 10, 2012

Report

Report Number
MW5024247
Event Type
Death
Date Received
February 10, 2012
Date of Event
January 31, 2012
Report Date
January 31, 2012
Manufacturer
DRAEGER MEDICAL INC
Product Code
MHX
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
PA, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

RN OBSERVED PT O2 SATS IN THE LOW 60'S AND CALLED A CODE. UPON OBSERVING ALL THE ALARM MONITORS, ALL THE ALARMS WERE OFF EXCEPT HEART RATE. HIT THE BUTTON TO RESET THE ALARMS TO THE ICU STANDARDS AND THE ALARMS REMAINED OFF. REPORTED TO RN SUPERVISOR AND ICU SUPERVISOR/MGR MANUALLY TURNED ALARMS ON. PT WAS IN PEA WITH HEART RATE IN 60'S TO 70'S SO THE HEART RATE ALARM DID NOT ALARM. THERE WAS NO VISUAL DISPLAY THAT ALARMS WERE "OFF."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONITOR DOCKING STATION MHX DRAEGER MEDICAL INC 12095054

Patients

Seq Age Sex Outcome Treatment
1 92 YR Death