FDA Adverse Event Injury Summary report: N

BLOM-SINGER DUAL VALVE INDWELLING VOICE PROSTHESIS

MDR report key: 24604342 · Received March 16, 2026

Report

Report Number
2025182-2026-00001
Event Type
Injury
Date Received
March 16, 2026
Date of Event
November 3, 2025
Report Date
March 11, 2026
Manufacturer
FREUDENBERG MEDICAL, LLC
Product Code
EWL
UDI-DI
00813307010951
PMA / PMN Number
K082206
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPON VISUAL INSPECTION OF THE VOICE PROSTHESIS IT WAS CLEAR THAT THE FLAP ON THE ESOPHAGEAL SIDE OF THE DEVICE WAS MISSING. THERE ARE NO OBVIOUS TEARS RIPS OR ABRASIONS VISIBLE UNDER THE MICROSCOPE, IN THE BARREL OF THE VOICE PROSTHESIS OR AROUND WHERE THE ESOPHAGEAL VALVE FLAP WOULD HAVE BEEN LOCATED. THE VALVE FLAP THAT WAS ASPIRATED AND FOUND IN THE KARINA VIA TRACHEOSCOPY, WAS NOT PROVIDED FOR INSPECTION.

Description of Event or Problem · 0

THE SPEAKING VALVE WAS INSERTED 5 DAYS PRIOR TO THE INCIDENT. THE PATIENT ATTENDED ED WITH UNCONTROLLED ASPIRATION OF SECRETIONS AND FOOD/FLUIDS FROM VALVE. ON CLOSER INSPECTION OF THE VALVE, THERE APPEARED TO BE NO ANTERIOR FLAP. TRACHEOSCOPY IDENTIFIED A SMALL PLASTIC FLAP IN THE CARINA BUT WAS NOT SEEN ON XRAY. THE PATIENT REPORTED THAT HE HAD BEEN CLEANING HIS VALVE IN THE NORMAL WAY, WITH NO FORCE AND WITH THE CORRECT EQUIPMENT. SLT THAT PLACED THE VALVE SAID THAT SHE HAD NOT NOTICED ANY PROBLEMS WITH THE VALVE AT THE TIME OF INSERTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
697 BLOM-SINGER DUAL VALVE INDWELLING VOICE PROSTHESIS BLOM-SINGER DUAL VALVE INDWELLING VOICE PROSTHESIS EWL FREUDENBERG MEDICAL, LLC DV 2010-NS 0000321626 00813307010951

Patients

Seq Age Sex Outcome Treatment
1 NA Male Other