THERMOCOOL SMARTTOUCH SF
Report
- Report Number
- 2029046-2026-00850
- Event Type
- Malfunction
- Date Received
- March 14, 2026
- Date of Event
- January 12, 2026
- Manufacturer
- BIOSENSE WEBSTER INC
- Product Code
- LPB
- UDI-DI
- 10846835009804
- PMA / PMN Number
- P030031/S078
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION DETAILS: THE DEVICE WAS RETURNED TO JOHNSON & JOHNSON MEDTECH (J&J MEDTECH) FOR EVALUATION. FOLLOWING J&J MEDTECH PROCEDURES, A VISUAL INSPECTION AND MAGNETIC TEST OF THE RETURNED DEVICE WERE PERFORMED. VISUAL INSPECTION REVEALED THAT A REDDISH MATERIAL, PRESUMABLY BLOOD, WAS OBSERVED INSIDE THE PEBAX. THE DEVICE WAS CONNECTED TO THE CARTO 3 SYSTEM AND IT WAS RECOGNIZED AND VISUALIZED CORRECTLY. NO MAGNETIC ISSUES OR ERRORS WERE OBSERVED. IN A CLOSER INSPECTION A HOLE WAS OBSERVED ON THE SURFACE OF THE PEBAX. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH NUMBER, AND NO INTERNAL ACTIONS WERE IDENTIFIED. THE REDDISH MATERIAL INSIDE THE PEBAX COULD BE RELATED TO THE ICON JUMPING REPORTED BY THE CUSTOMER; THEREFORE, THE CUSTOMER COMPLAINT WAS CONFIRMED. THE POTENTIAL CAUSE OF THE PEBAX DAMAGE COULD BE RELATED TO THE MANIPULATION OF THE DEVICE DURING THE PROCEDURE; HOWEVER, THIS CANNOT BE CONCLUSIVELY DETERMINED. THE INSTRUCTIONS FOR USE (IFU) CONTAIN THE FOLLOWING RECOMMENDATIONS: WHEN CLEANING THE TIP ELECTRODE, BE CAREFUL NOT TO TWIST THE TIP ELECTRODE WITH RESPECT TO THE CATHETER SHAFT; TWISTING MAY DAMAGE THE TIP ELECTRODE BOND AND LOOSEN THE TIP ELECTRODE, OR MAY DAMAGE THE CONTACT FORCE SENSOR. A SIGNIFICANT CHANGE IN THE BASELINE READING AFTER CLEANING MIGHT INDICATE A DAMAGED CONTACT FORCE SENSOR. IN ADDITION, IN ORDER TO PREVENT DAMAGE TO THE CATHETER TIP, VERIFY COMPATIBILITY BETWEEN THE SHEATH AND CATHETER, ADVANCE THE CATHETER THROUGH THE SHEATH PRIOR TO INSERTION. ANY SHEATH < 8.5 F IS CONTRAINDICATED. AS PART OF JOHNSON & JOHNSON MEDTECH¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NOTE: THE ¿SUSPECTED MEDICAL DEVICE¿ REPORTED IN SECTION D OF THIS REPORT IS NOT MARKETED IN USA OR APPROVED BY THE FDA. HOWEVER, IT IS BEING REPORTED AS BIOSENSE WEBSTER CONSIDERS THIS AS A SIMILAR DEVICE TO THERMOCOOL SMARTTOUCH SF APPROVED UNDER P030031/S078. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY BIOSENSE WEBSTER INC., OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, BIOSENSE WEBSTER INC., OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MANUFACTURER'S REF. # (B)(4).
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A THERMOCOOL SMARTTOUCH SF FOR WHICH BIOSENSE WEBSTER¿S PRODUCT ANALYSIS LAB (PAL) IDENTIFIED A HOLE ON THE PEBAX SUFACE. IT WAS INITIALLY REPORTED BY THE CUSTOMER THAT DURING THE OPERATION, THE ICON WAS WAGGLING ON THE CARTO SYSTEM SCREEN. A SECOND DEVICE WAS USED TO COMPLETE THE OPERATION. THERE WAS NO ADVERSE EVENT REPORTED ON PATIENT. THE CUSTOMER'S REPORTED WAGGLING ICON ISSUE IS NOT CONSIDERED TO BE MDR REPORTABLE SINCE THE POTENTIAL RISK THAT IT COULD CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR DEATH TO THE OPERATOR OR PATIENT IS REMOTE. ON 20-FEB-2026, THE BWI PAL REVEALED THAT A VISUAL INSPECTION OF THE RETURNED DEVICE FOUND A HOLE IN THE PEBAX AND A REDDISH MATERIAL INSIDE THE PEBAX. THESE FINDINGS WERE REVIEWED AND ASSESSED THE ISSUE OF A ¿HOLE¿ IN THE PEBAX AS AN MDR REPORTABLE MALFUNCTION SINCE THE INTEGRITY OF THE DEVICE HAS BEEN COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 660980 | THERMOCOOL SMARTTOUCH SF | CARDIAC ABLATION PERCUTANEOUS CATHETER | LPB | BIOSENSE WEBSTER INC | 31685026L | 10846835009804 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |