FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 24598912 · Received March 13, 2026

Report

Report Number
1645337-2026-02840
Event Type
Injury
Date Received
March 13, 2026
Date of Event
February 5, 2026
Manufacturer
MENTOR TEXAS
Product Code
FTR
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION REVIEW COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: IMPLANT EXCHANGE. D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

THE MENTOR FAILURE ANALYSIS LAB RECEIVED TWO DEVICES FOR EVALUATION ON 01-APR-2026 (CATALOG #3503251BC, LOT #5851961 & 5842858. ON 17-APR-2026, MENTOR BECAME AWARE THAT THE DEVICE FROM LOT #5851961 IS THE COMPLAINT DEVICE. BLOCK D HAS BEEN UPDATED ACCORDINGLY. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH AN UNSPECIFIED MENTOR GEL BREAST PROSTHESIS THAT RUPTURED POST IMPLANTATION. RUPTURE OF THE PATIENT¿S RIGHT BREAST PROSTHESIS WAS DISCOVERED DURING AN IMPLANT EXCHANGE SURGERY. THE PATIENT¿S EXPLANTED BREAST PROSTHESES WERE REPLACED WITH NON-MENTOR BREAST PROSTHESES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
158419 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 5851961

Patients

Seq Age Sex Outcome Treatment
1