FDA Adverse Event Malfunction Summary report: N

DA VINCI XI

MDR report key: 24580384 · Received March 12, 2026

Report

Report Number
2955842-2026-15797
Event Type
Malfunction
Date Received
March 12, 2026
Date of Event
February 19, 2026
Report Date
March 11, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874115343
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE WENT ON-SITE AND REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (IESU). THE SYSTEM WAS VERIFIED AND READY FOR USE. INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE DA VINCI PRODUCT WITH AN ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/OR IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED ENDOMETRIOSIS RESECTION SURGICAL PROCEDURE, USER INFORMED THE TECHNICAL SUPPORT ENGINEER (TSE) THAT AN ERROR C-00 OCCURRED ON THE ERBE GENERATOR DURING SYSTEM POWER-ON SELF-TEST. INITIALLY, THE USER ATTEMPTED TO RESOLVE THE ISSUE BY POWER CYCLING THE GENERATOR, BUT THE ERROR PERSISTED. THE TSE TRIED TO REVIEW ERROR LOGS, BUT THE ONSITE CONNECTION WAS NOT POSTING DUE TO INTERNET ISSUES IN THE THEATER. LATER, USER CALLED BACK AND REPORTED THAT THE ERROR C-00 WAS PERMANENT, AND TROUBLESHOOTING STEPS GUIDED VIA VIDEO DID NOT RESOLVE THE ISSUE. THE USER HAD A THIRD-PARTY ENERGY ACTIVATION CABLE BUT ONLY AN FT10, WHICH IS NOT VALIDATED FOR USE WITH THE DA VINCI SYSTEM. CONSEQUENTLY, THE USER DECIDED TO START THE CASE LAPAROSCOPICALLY, AND THE SYSTEM WAS MARKED AS DOWN. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
180 DA VINCI XI VISION SIDE SYSTEM, RECONDITIONED NAY INTUITIVE SURGICAL, INC 380721-10 N/A 00886874115343

Patients

Seq Age Sex Outcome Treatment
1