SPRINT QUATTRO
Report
- Report Number
- 2649622-2012-01973
- Event Type
- Malfunction
- Date Received
- February 16, 2012
- Date of Event
- December 1, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S17
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA THAT COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ONE - PATIENT ALERT FOR LEAD FAILURE PREDICTOR (LFP) ON (B)(4) 2011 13:57:12. FOUR - LFP HIGH RATE-NS <= 205 MS AVERAGE V-CYCLE ON (B)(4) 2011 IN THE TIMEFRAME BETWEEN 12:47:17 AND 14:27:42. VENTRICULAR SHORT INTERVAL COUNT V-SIC=2.6 COUNTS AVG/DAY, IN (B)(4), BETWEEN (B)(4) 2011 12:15:22 AND (B)(4) 2011 11:39:37.
IT WAS REPORTED THAT ONE DAY POST-IMPLANT, NST (NON-SUSTAINED TACHYCARDIA) EPISODES WERE RECORDED. THE DEVICE AND LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6944 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Other | 5594 IMPLANTABLE PACING LEAD| 4396 IMPLANTABLE PACING LEAD |