SABINA MOBILE LIFT
Report
- Report Number
- 8030916-2026-00026
- Event Type
- Malfunction
- Date Received
- March 11, 2026
- Date of Event
- February 13, 2026
- Report Date
- May 20, 2026
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FSA
- UDI-DI
- 00887761981782
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
NO FURTHER INFORMATION IS AVAILABLE ON THE REPAIR OF THE DEVICE AT THIS TIME. THE INVESTIGATION IS ONGOING, HOWEVER IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED FOLLOWING THE COMPLETION OF THE INVESTIGATION, THE ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED FOLLOWING COMPLETION OF THE INVESTIGATION, THE ADDITIONAL RELEVANT INFORMATION WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT. THE SABINA SIT-TO-STAND LIFT IS ESPECIALLY DESIGNED FOR PEOPLE WHO HAVE DIFFICULTY IN STANDING UP ON THEIR OWN FROM A SEATED POSITION. SABINA SIT-TO-STAND LIFT IS INTENDED FOR USE WITH PATIENTS WHO ARE ABLE TO ACTIVELY PARTICIPATE IN THE RAISING MOTION. WHEN STANDING, THEY CAN BE MOVED TO A WHEELCHAIR OR TO A TOILET, THIS GIVES THEM STANDING PRACTICE IN CONNECTION WITH THE TRANSFER. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS DEVICE. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR DEVICE.
THE BAXTER TECHNICIAN FOUND THE MAST WELDMENT NEEDED TO BE REPLACED. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS DEVICE. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR DEVICE. THE SABINA SIT-TO-STAND LIFT IS ESPECIALLY DESIGNED FOR PEOPLE WHO HAVE DIFFICULTY IN STANDING UP ON THEIR OWN FROM A SEATED POSITION. SABINA SIT-TO-STAND LIFT IS INTENDED FOR USE WITH PATIENTS WHO ARE ABLE TO ACTIVELY PARTICIPATE IN THE RAISING MOTION. WHEN STANDING, THEY CAN BE MOVED TO A WHEELCHAIR OR TO A TOILET, THIS GIVES THEM STANDING PRACTICE IN CONNECTION WITH THE TRANSFER. THE TECHNICIAN REPLACED THE MAST WELDMENT TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED. ALTHOUGH THERE WAS NO REPORTED INJURY WITH THIS EVENT, IF THE REPORT OF A CRACKED MAST WERE TO RECUR, IT COULD POTENTIALLY CAUSE SERIOUS INJURY OR DEATH. THEREFORE, BAXTER IS REPORTING THIS EVENT.
ON 13-FEB-2026, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT SABINA II EE BASIC (PRODUCT CODE P2020003, SERIAL NUMBER (B)(6)), HAD MAST WELDMENT CRACKED. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 476296 | SABINA MOBILE LIFT | LIFT, PATIENT, NON-AC-POWERED | FSA | BAXTER HEALTHCARE CORPORATION | P2020003 | 00887761981782 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | N/A. |