FDA Adverse Event Malfunction Summary report: N

ANTERALIGN¿ SPINAL SYSTEM WITH TITAN NANOLOCK¿ SURFACE TECHNOLOGY

MDR report key: 24566271 · Received March 11, 2026

Report

Report Number
3003120897-2026-00403
Event Type
Malfunction
Date Received
March 11, 2026
Date of Event
April 18, 2025
Report Date
March 11, 2026
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC.
Product Code
MAX
UDI-DI
00763000572174
PMA / PMN Number
K212524
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS OF PART# 4680004, LOT# K23F1290 VISUAL AND FUNCTIONAL TESTING IDENTIFIED THAT THE INNER SHAFT OF THE INSTRUMENT IS MISSING. THE INSTRUMENT CANNOT FUNCTION AS INTENDED. THIS REGULATORY REPORT IS BEING SUBMITTED DUE TO RETROSPECTIVE REVIEW. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE INSERTER WOULD NO LONGER THREAD INTO THE DEVICE. THERE WAS NO PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
628297 ANTERALIGN¿ SPINAL SYSTEM WITH TITAN NANOLOCK¿ SURFACE TECHNOLOGY INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, MAX MEDTRONIC SOFAMOR DANEK USA, INC. 4680004 K23F1290 00763000572174

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown