FDA Adverse Event Malfunction Summary report: N

ATTAIN STARFIX

MDR report key: 2456585 · Received February 16, 2012

Report

Report Number
2649622-2012-01790
Event Type
Malfunction
Date Received
February 16, 2012
Date of Event
December 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P060039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. IMPEDANCE - HIGH RESISTANCE/IMPEDANCE: 2 - PATIENT ALERTS FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(6) 2011 09:00:12 AND (B)(6) 2011 03:00:12. PROGRAMMER S2D DATA SHOW ALERTS FOR LV PACING LEAD IMPEDANCE > 3000 OHMS ON (B)(6) 2011 09:00:12 AND (B)(6) 2011 03:00:12. IMPEDANCE - LOW RESISTANCE/IMPEDANCE: 1 - PATIENT ALERT FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(6) 2011 15:00:04. PROGRAMMER S2D DATA SHOWS AN ALERT FOR RV DEFIB LEAD IMPEDANCE= 19 OHMS ON (B)(6) 2011 15:00:04. IMPEDANCE - RESISTANCE/IMPEDANCE INCREASE: WEEKLY HV-LEAD IMPEDANCE TREND DATA SHOWS AN INCREASE FOR MAX DEFIB ACTIVE CAN ON AND MAX SVC DEFIB IMPEDANCE= 24 TO 110 OHMS PEAK BETWEEN (B)(6) 2011. SENSING - OVERSENSING: 1 - VF=190 MS AVERAGE V-CYCLE ON (B)(6) 2011 15:58:37. SENSING - HIGH RESISTANCE/IMPEDANCE: 1 - PATIENT ALERT FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(6) 2011 03:00:14. WEEKLY PACE LEAD IMPEDANCE TREND DATA SHOWS A SPIKE INCREASE FOR MAX VENTRICULAR PACE BI IMPEDANCE = 798 TO 4047 OHMS PEAK BETWEEN (B)(6) 2011, THEN RETURNING TO A BASELINE OF 342 OHMS ON (B)(6) 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD ALERT TRIGGERED, AND THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH IMPEDANCES. SHORTLY AFTER THAT, THE RIGHT VENTRICULAR (RV) AND SUPERIOR VENA CAVA (SVC) COILS EXHIBITED HIGH IMPEDANCES. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN STARFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4195

Patients

Seq Age Sex Outcome Treatment
1 61 YR Other 5076 IMPLANTABLE PACING LEAD| 511211 X2 COMPETITOR IMPLANTABLE PACING LEADS| D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB