FDA Adverse Event Malfunction Summary report: N

INTELLIVUE MULTI MEASUREMENT SERVER X2

MDR report key: 24560487 · Received March 10, 2026

Report

Report Number
9610816-2026-100478
Event Type
Malfunction
Date Received
March 10, 2026
Date of Event
February 10, 2026
Report Date
March 10, 2026
Manufacturer
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
Product Code
MHX
UDI-DI
00884838000261
PMA / PMN Number
K150310
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A PHILIPS REMOTE SERVICE ENGINEER (RSE) SPOKE TO THE CUSTOMER WHO INDICATING THAT THE DOWN POST ON THE GCX ARM BROKE OFF AND THE X2 FELL OFF AND HIT THE FLOOR. ON FURTHER INVESTIGATION THE RSE WAS ADVISED THAT THE DOWN POST ON THE ARM HAD BECOME LOOSE OVER THE YEARS AND WAS BEING HELD ON BY ONLY A ONE OR TWO SCREWS. THE SCREWS SHEARED OFF AND THE X2 FELL. IT WAS UNKNOWN WHO ORIGINALLY INSTALLED THE GCX ARM. BASED ON THE RESULTS OF THE ANALYSIS, IT WAS DETERMINED THAT THE PRODUCT HAS MALFUNCTIONED AND THE MOST LIKELY CAUSE OF THE REPORTED PROBLEM WAS A MAINTENANCE ISSUE. THE CUSTOMER REMOUNTED THE GCX ARM TO RESOLVE THE ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE GCX ARM BROKE OFF AND THE X2 FELL OFF AND HIT THE FLOOR. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621346 INTELLIVUE MULTI MEASUREMENT SERVER X2 MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) MHX PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH 865039 00884838000261

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown