FDA Adverse Event Malfunction Summary report: N

LUER-LOK

MDR report key: 24545742 · Received March 9, 2026

Report

Report Number
1911916-2026-00106
Event Type
Malfunction
Date Received
March 9, 2026
Date of Event
February 4, 2026
Report Date
March 12, 2026
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
00382903096534
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP FOR DEVICE EVALUATION. MOISTURE WAS REPORTED IN SYRINGES. TO SUPPORT THE INVESTIGATION, THREE SAMPLES IN OPENED BLISTER PACKAGING WERE RECEIVED AND EVALUATED BY THE QUALITY TEAM. VISUAL INSPECTION IDENTIFIED SILICONE LUBRICANT AT THE BOTTOM OF THE SYRINGE BARREL, WHICH IS EXPECTED AND ACCEPTABLE. THE MEDICAL GRADE SILICONE IS INTENTIONALLY APPLIED TO THE INNER BARREL WALL AND RUBBER STOPPER TO FACILITATE SMOOTH PLUNGER MOVEMENT. A DEVICE HISTORY RECORD REVIEW FOR MATERIAL NUMBER 309653, LOT 5142418, IDENTIFIED NO NONCONFORMANCES DURING THE MANUFACTURING OF THIS LOT. NO RELATED QUALITY NOTIFICATIONS WERE FOUND, AND ALL PROCESSES AND FINAL INSPECTIONS MET SPECIFICATION REQUIREMENTS. BASED ON THE INVESTIGATION AND ANALYSIS OF THE RETURNED SAMPLES, THE CUSTOMERS REPORTED CONDITION WAS CONFIRMED, HOWEVER, THE OBSERVATION CORRESPONDS TO THE PRODUCTS SILICONE LUBRICATION AND IS CONSIDERED ACCEPTABLE.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD.

Description of Event or Problem · 0

IT IS REPORTED FOREIGN MATTER. DESCRIPTION: WE HAVE RECENTLY DISCOVERED MOISTURE IN SEVERAL OF OUR 50 ML LUER-LOCK SYRINGES THAT ARE STORED IN OUR SUPPLY ROOM IN THEIR ORIGINAL PACKAGING AND NOT EXPOSED TO ANY MOISTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35030 LUER-LOK SYRINGE, PISTON FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 5142418 00382903096534

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown