FDA Adverse Event
Malfunction
Summary report: N
QUANTUM ROLLER PUMP - 6"
MDR report key: 24543685
·
Received March 9, 2026
Report
- Report Number
- 3006073153-2026-00058
- Event Type
- Malfunction
- Date Received
- March 9, 2026
- Date of Event
- August 21, 2023
- Report Date
- March 9, 2026
- Manufacturer
- SPECTRUM MEDICAL LTD
- Product Code
- DTQ
- UDI-DI
- 05060434420787
- PMA / PMN Number
- K173834
- Removal / Correction Number
- Z-0224-2025
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
ISSUE WAS CONFIRMED UPON ARRIVAL. THIS EVENT WAS ORIGINALLY REPORTED IN A SUMMARY REPORT 3006073153-2025-00059, WHICH WAS SUBMITTED IN RELATION TO THE AFFECTED DEVICES FROM THE VOLUNTARY RECALLED PERFORMED BY SPECTRUM MEDICAL LTD (RECALL NUMBER Z-0224-2025). FDA HAS SINCE REQUESTED THAT EACH EVENT FOR THE DEVICES RELATED TO THE VOLUNTARY RECALL BE INDIVIDUALLY SUBMITTED, RESULTING IN THE CREATION AND SUBMISSION OF THIS REPORT.
Description of Event or Problem · 0
USER REPORTED THAT THE BOBBINS NO LONGER LOCK WHEN THE FLAG IS IN THE CLOSED POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2840 | QUANTUM ROLLER PUMP - 6" | ROLLER PUMP | DTQ | SPECTRUM MEDICAL LTD | QRP6 | 05060434420787 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |