I-STAT ACT KAOLIN CARTRIDGE
Report
- Report Number
- 2245578-2026-00046
- Event Type
- Malfunction
- Date Received
- March 7, 2026
- Date of Event
- January 5, 2026
- Report Date
- May 20, 2026
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- JBP
- PMA / PMN Number
- K023582
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT#: (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON 17-FEB-2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT ACT KAOLIN CARTRIDGES THAT YIELDED A DISCREPANT RESULTS ON A 73-YEAR-OLD MALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE. DATE: TESTED: RESULT: HEPARIN: HEPARIN TIME: ON (B)(6) 2026, 20000, 8:26, ON (B)(6) 2026, 8:38, 322, ON (B)(6) 2026, 3000, 8:40, ON (B)(6) 2026, 8:51, 332, ON (B)(6) 2026, 4000, 8:52, ON (B)(6) 2026, 9:04, 332, ON (B)(6) 2026, 4000, 9:08, ON (B)(6) 2026, 9:19, 353, ON (B)(6) 2026, 2000, 9:23, ON (B)(6) 2026, 9:36, 363, ON (B)(6) 2026, 10:05, 173, ON (B)(6) 2026, 10:21, 414, ON (B)(6) 2026, 10:43, 378, ON (B)(6) 2026, 11:02, 373, ON (B)(6) 2026, 11:37, 194. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. AN INVESTIGATION IS UNDERWAY. PER I-STAT SYSTEM MANUAL (ART: 714185-00Q), THE I-STAT KAOLIN ACTIVATED CLOTTING TIME (KAOLIN ACT) TEST IS AN IN VITRO DIAGNOSTIC TEST THAT USES FRESH, WHOLE BLOOD. AND IS USED TO MONITOR HIGH-DOSE HEPARIN ANTICOAGULATION FREQUENTLY ASSOCIATED WITH CARDIOVASCULAR SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 43344 | I-STAT ACT KAOLIN CARTRIDGE | ACT KAOLIN CARTRIDGE | JBP | ABBOTT POINT OF CARE INC. | NA | R25277 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Male |