FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2454011 · Received February 15, 2012

Report

Report Number
2531779-2012-01161
Event Type
Malfunction
Date Received
February 15, 2012
Report Date
October 25, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). CORRECTION NUMBER: 2531779-03/24/2010-003-R. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 01/19/2012. THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE REWIND, LOAD, AND PRIME SEQUENCE WAS SUCCESSFULLY COMPLETED. NO LOSS OF PRIME WARNINGS WERE OBSERVED IN THE HISTORY AND THE COMPLAINT COULD NOT BE DUPLICATED. EVALUATION REVEALED THAT THE FORCE SENSOR WAS OUT OF CALIBRATION AND THE FORCE RESISTANCE MEASUREMENT WAS OUT OF SPECIFICATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED THAT THE FORCE SENSOR WAS OUT OF CALIBRATION AND THE FORCE RESISTANCE MEASUREMENT WAS OUT OF SPECIFICATION. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 01/19/2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 19 YR