Report Number
1818910-2026-04375
Event Type
Malfunction
Date Received
March 6, 2026
Date of Event
February 25, 2026
Manufacturer
DEPUY IRELAND - 3015516266
Product Code
LXH
UDI-DI
10603295129752
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, ¿ANKLE CLAMP JAMMED¿ ADDITIONALLY, "HI, IT WAS 2 PRODUCTS, THE ATTUNE EM TIBIAL DIST UPROD AND THE ATTUNE EM TIBIAL ANKLE CLAMP. THESE ITEMS WERE JAMMED TOGETHER". THE PRODUCT WAS NOT RETURNED TO DEPUY SYNTHES, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. REVIEW OF THE PHOTOGRAPHIC EVIDENCE REVEALED THAT THE ATTUNE EM TIBIAL DIST UPROD WAS ASSEMBLED WITH ITS MATING DEVICE. THE JAMMED CONDITION CANNOT BE CONFIRMED BASED ON THE OBSERVED CONDITION IN THE ATTACHMENTS, AS FUNCTIONALITY ISSUES CANNOT BE ASSESSED THROUGH A PHOTOGRAPHIC INVESTIGATION. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE OBSERVED CONDITION OF THE ATTUNE EM TIBIAL DIST UPROD WOULD NOT CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: ADDED: B5. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED: CAN YOU PLEASE CONFIRM THE QUANTITY OF PRODUCTS INVOLVED? IT WAS 2 PRODUCTS, THE ATTUNE EM TIBIAL DIST UPROD AND THE ATTUNE EM TIBIAL ANKLE CLAMP. THESE ITEMS WERE JAMMED TOGETHER.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE ANKLE CLAMP IS JAMMED. THERE WAS NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
549999 ATTUNE ATTUNE INSTRUMENTS : ALIGNMENT DEVICES LXH DEPUY IRELAND - 3015516266 PG352249 10603295129752

Patients

Seq Age Sex Outcome Treatment
1 0 DA Unknown